Specific ranges of the Analog View are:
Ref.
Feature
Color
Description
1
Normal
Range
White
Area of the display range where an alarm will not
be triggered.
2
Caution
Range
Yellow
Area of the display range that provides a caution
indicator.
3
Alarming
Range
Red
Area of the display range where an alarm will be
triggered.
Some ranges display as quarter circles, others display as half circles. A quarter
circle displays when the value has a physiologic normal level at one end of the
range. A half circle displays when the value has a physiologic normal level in
the middle of the display range.
In the example below, the SpO
2
gauge is shown as a quarter circle, where
values lower than 88% will trigger an alarm, and the PR gauge is shown as a
half circle, where values below 50 bpm and above 140 bpm will trigger an
alarm.