Root NIBPT Safety Information, Warnings, and Cautions
www.masimo.com 29 Masimo
CAUTION: An operator may only perform maintenance procedures
specifically described in the manual. Refer servicing to qualified service
personnel trained in the repair of this equipment.
CAUTION: Electrical Shock Hazard: Carry out periodic tests to verify that
leakage currents of patient-applied circuits and the system are within
acceptable limits as specified by the applicable safety standards. The
summation of leakage currents must be checked and in compliance with IEC
60601-1 and UL60601-1. The system leakage current must be checked when
connecting external equipment to the system. When an event such as a
component drop of approximately 1 meter or greater or a spillage of blood
or other liquids occurs, retest before further use. Injury to personnel could
occur.
Note: Excessive cleaning solution can flow into the device and cause
damage to internal components.
Compliance Warnings and Cautions
WARNING: Changes or modifications not expressly approved by Masimo
shall void the warranty for this equipment and could void the user’s
authority to operate the equipment.
WARNING: Do not incinerate the battery.
WARNING: In accordance with international telecommunication
requirements, the frequency band of 2.4 GHz and 5.15 to 5.25 GHz is only
for indoor usage to reduce potential for harmful interference to co-channel
mobile satellite systems.
WARNING: Users are advised that high-power radars are allocated as
primary users (i.e. priority users) of the bands 5.25-5.35 GHz and 5.65-5.85