Root NIBPT Safety Information, Warnings, and Cautions
www.masimo.com 28 Masimo
CAUTION: Kite is not a primary display. Medical decisions should be made
using data from the primary display of a device in conjunction with clinical
signs and symptoms.
Patient SafetyNet System
Note: The wireless communication status between Root and Patient
SafetyNet is displayed by Patient SafetyNet.
Cleaning and Service Warnings and Cautions
WARNING: Electrical Shock Hazard: The Root battery should be installed
and/or removed from Root only by qualified personnel.
WARNING: Do not use petroleum-based or acetone solutions, or other
harsh solvents, to clean the Root. These substances affect the device’s
materials and device failure can result.
CAUTION: Do not touch, press, or rub the display panels with abrasive
cleaning compounds, instruments, brushes, rough-surface materials, or
bring them into contact with anything that could scratch the display.
CAUTION: Do not submerge the Root in any cleaning solution or attempt to
sterilize by autoclave, irradiation, steam, gas, ethylene oxide or any other
method. This will seriously damage the device.
CAUTION: To prevent damage to the temperature probe do not autoclave.
CAUTION: Electrical shock and flammability hazard: Before cleaning, always
turn off the device and disconnect from any AC power source.