Root NIBPT Safety Information, Warnings, and Cautions
www.masimo.com 27 Masimo
Optional Integrated Temperature
WARNING: The use of any other probe cover may produce temperature
measurement errors or may result in inaccurate readings.
WARNING: Do not take axillary temperature through the patient's clothing,
direct probe-cover-to-skin contact is required.
CAUTION: Long-term continuous temperature monitoring, greater than 5
minutes, is not recommended.
CAUTION: Ensure probe well is properly in place.
CAUTION: Biting the probe tip while taking a temperature may result in
damage to the probe.
CAUTION: Use of the incorrect probe at the measurement site will result in
temperature errors.
Note: If a reliable spot check measurement cannot be made, the
temperature module will automatically switch to continuous mode to make
the measurement.
Note: Verify the compatibility of the temperature probe and probe cover
before use with Root.
Kite
WARNING: When using Root the Kite accessory does not generate or
manage alarms. The Root alarms, used in conjunction with clinical signs and
symptoms, are the primary sources for determining that an alarm condition
exists.