Operator's Manual
Root NIBPT Safety Information, Warnings, and Cautions
www.masimo.com 26 Masimo
CAUTION: The wireless quality of services may be influenced by the
presence of other devices that may create radio frequency interference (RFI).
Some RFI devices to consider are as follows: electrocautery equipment,
cellular telephones, wireless PC and tablets, pagers, RFID, MRI, electrically
powered wheelchair, etc. When used in the presence of potential RFI
devices, consideration should be taken to maximize separation distances
and to observe for any potential signs of interference such as loss of
communication or reduced Wi-Fi signal strength.
Note: Root is provided with a Wi-Fi signal indicator as an indication of Wi-Fi
communication.
Note: Root’s alarm capabilities have been designed to be independent of
the Wi-Fi communication feature in order to preserve Root’s primary alarms.
Note: When the Root monitor is in synchronized waveform view, the EEG
Chart Speed of the SedLine Window cannot be changed.
Noninvasive Blood Pressure
WARNING: Before applying the cuff on the patient, confirm the cuff size is
appropriate.
WARNING: When a blood pressure measurement error code occurs, any
blood pressure values reported should be disregarded.
CAUTION: If the blood pressure cuff is on the same limb as monitoring
equipment (i.e., pulse oximeter probe), the pressurization within the cuff can
cause temporary loss of function of the monitoring equipment.