Root NIBPT Safety Information, Warnings, and Cautions
www.masimo.com 25 Masimo
CAUTION: Before using Root under high intensity surgical lights, confirm
that the display settings allow for clear display of measurements.
CAUTION: Do not connect to an electrical outlet controlled by a wall switch
or dimmer.
CAUTION: Do not place the Root on electrical equipment that may affect
the device, preventing it from working properly.
CAUTION: Failure to charge Root promptly after a Low Battery alarm may
result in the device shutting down.
CAUTION: To minimize radio interference, other electrical equipment that
emits radio frequency transmissions should not be in close proximity to
Root.
CAUTION: If the Radical-7 or Radius-7 stops communicating with Root,
parameters and measurements will not show on the Root; however, this will
not affect Radical-7's or Radius-7's ability to monitor the patient.
CAUTION: In order to establish and maintain Root’s minimum Quality of
Service, the following network specifications should be met before and after
installation:
Wired Network Connection
During Ping Test, passing result if:
a. At least 98% of packets have latency ≤ 30 milliseconds, and
b. No more than 2 % packets loss.
Wireless Network Connection
During Ping Test, passing result if:
a. At least 98% of packets have latency ≤ 100 milliseconds,
b. No more than 2 % packets loss, and
c. Primary access point signal strength at least -67 dBm.