Operator's Manual
Root NIBPT Chapter 16: Service and Maintenance
www.masimo.com 244 Masimo
11. Disconnect the manometer, 500mL bottle or cuff wrapped around
solid object, and hand bulb from the NIBP module.
12. Power OFF the manometer.
Test Method for Neonates
1. Follow steps 1 through 3 from
Test Method for Adults and
Pediatrics
above.
Note: If using a cuff, verify the cuff is Neonate. Do not use an Adult
or Pediatric cuff for this test method.
2. On the Root Main Screen, touch the
Profiles
option in the Status
Bar. See
About the Status Bar
on page 53.
3. Select Neonates in the
Profiles
screen, then touch OK.
4. Start the NIBP measurement then immediately abort:
Touch the Start NIBP button.
Touch the Start NIBP button again when then pump starts to
inflate the system.
5. Press the Main Menu icon
6. Press NIBP .
7. Press CALIBRATION .
For the password, enter 4258.
Press the Enter key.
8. Press OVERPRESSURE TEST .
9. To begin the overpressure test, press the Test button, then
increase the pressure to the overpressure point: