Operator's Manual

Root NIBPT Chapter 16: Service and Maintenance
www.masimo.com 241 Masimo
Adults, Pediatrics
300 ±10mmHg
Neonates
150 ±5mmHg
Equipment
Calibrated Manometer*
Pneumatic "T" Adapters
500mL bottle or regular-sized cuff wrapped around solid object
Hand Bulb
Connection Tubing
*Verify the manometer has been calibrated within the last 12 months.
Calibration accuracy shall be within ±0.02% to ±0.25% FS (Full Scale) and a
measurement uncertainty of ±1 LSD (Least Significant Digit).