Operator's Manual
Root NIBPT Chapter 16: Service and Maintenance
www.masimo.com 240 Masimo
For the password, enter 4258.
Press the Enter key.
7. Press AIR LEAK TEST .
8. To begin the air leak test, press Test.
Wait for the countdown timer to reach 0 second.
At the bottom of the screen, in the Result section, verify the
leak rate is less than 6mmHg. If the leak rate is greater than
6mmHg, verify there are no leaks in the calibration equipment
(tubing, 500mL container, arm cuff, etc.). If there are no leaks
detected from the calibration equipment and the leak rate is
still greater than 6mmHg, contact Masimo Technical Services.
See
Masimo Technical Services
on page 248.
9. Disconnect the manometer, 500mL bottle, and hand bulb from the
NIBP module.
10. Power OFF the manometer.
Overpressure Test
Note: This section is provided as a reference and intended for authorized
service personnel only.
Pass Criteria
International standards for automated NIBP devices require that the
pressure must not exceed 300mmHg on adults and pediatric patients and
150mmHg on neonatal patients with a tolerance of 10% for 15 seconds or
greater than 10% for 3 seconds. The overpressure pass criteria for the
Advantage module are: