Operator's Manual
Root NIBPT Chapter 16: Service and Maintenance
www.masimo.com 239 Masimo
Procedure
1. Connect the manometer, 500mL bottle, and hand bulb to the NIBP
module using "T" adapters and connection tubing.
2. Power ON the manometer.
Set the unit of measurement to mmHg.
Verify the manometer has been calibrated.
3. Power ON the Root.
4. On the Root home screen, press the Main Menu icon .
5. Press NIBP .
6. Press CALIBRATION .