Operator's Manual

Root NIBPT Chapter 16: Service and Maintenance
www.masimo.com 237 Masimo
8. Apply various pressures (0mmHg to 280mmHg) to the NIBP
module with the hand bulb.
9. Compare the NIBP module pressure to the manometer pressure:
If the NIBP module pressure and manometer pressure differ
WITHIN the ±3mmHg tolerance, then the NIBP module has
PASSED the calibration test and no further action is required.
Go to step 10.
Note: For pass criteria, see the
NIBP Module Calibration Test
on page 231.
If the NIBP module pressure and manometer pressure still
differ BEYOND the ±3mmHg tolerance, contact Masimo
Technical Services. See
Masimo Technical Services
on page
248.
10. Disconnect the manometer, 500mL bottle or cuff wrapped around
solid object, and hand bulb from the NIBP module.
11. Power OFF the manometer.