Operator's Manual
Root NIBPT Chapter 16: Service and Maintenance
www.masimo.com 233 Masimo
Procedure
1. Connect the manometer, 500mL bottle or cuff wrapped around
solid object, and hand bulb to the NIBP module using "T" adapters
and connection tubing.
2. Power ON the manometer.
Verify the manometer has been zeroed.
Set the unit of measurement to mmHg.
Verify the manometer has been calibrated.
3. Power ON the Root.
4. On the Root home screen, press the Main Menu icon .