Root NIBPT Chapter 16: Service and Maintenance
www.masimo.com 231 Masimo
Power-On Self Test
To conduct a Power-On Self Test
1. Connect Root to AC power, and verify that the AC Power Indicator
is illuminated.
2. Power On Root. Within five (5) seconds, all available indicators will
illuminate, the device will emit a tone, and the Masimo logo will
display.
NIBP Module Calibration Test
Note: This section is provided as a reference and intended for authorized
service personnel only.
Pass Criteria
International standards for automated NIBP devices require that the
maximum static pressure accuracy shall be ± 3mmHg or 2% or the reading,
whichever is greater. This is a stringent requirement and all test equipment
must be in excellent working order to properly perform this test. It is
important to verify the calibration before changing it. Historical data has
shown that the transducers rarely need to be re-calibrated although we still
suggest that the calibration be verified annually.
Equipment
Calibrated Manometer*
Pneumatic "T" Adapters
500mL bottle or regular-sized cuff wrapped around solid object
Hand Bulb