Operator's Manual
www.masimo.com 19 Masimo
Safety Information, Warnings, and
Cautions
CAUTION: Root is to be operated by, or under the supervision of, qualified
personnel only. The manual, accessories, directions for use, all precautionary
information, and specifications should be read before use. Refer to
Operator's Manuals for ISA, Kite, Patient SafetyNet, Radical-7, Radius-7,
TIR-1 and SedLine for additional safety information, warnings, and cautions.
Safety Warnings and Cautions
WARNING: Do not use Root if it appears or is suspected to be damaged.
WARNING: Do not adjust, repair, open, disassemble or modify Root. Injury
to personnel or equipment damage could occur. Return Root for servicing.
WARNING: Do not use Root during or nearby magnetic resonance imaging
(MRI) or in an MRI environment.
WARNING: Do not place Root or accessories in any position that might
cause it to fall on the patient.
WARNING: To ensure safety, avoid stacking multiple devices or placing
anything on the device during operation.
WARNING: Do not use Root in the presence of flammable anesthetics or
other flammable substance in combination with air, oxygen-enriched
environments or nitrous oxide to avoid risk of explosion.
WARNING: To reduce the risk of explosion, only replace battery with
Masimo supplied parts.
WARNING: Do not start or operate the Root unless the setup was verified
to be correct.