Chapter 15: Specifications
This chapter contains specifications of Root.
For information on the specifications of Radical-7, Radius-7, MOC-9
modules, and standalone devices, see Directions for Use or Operator's
Manuals for these devices.
Measurement Accuracy
NIBP
Pressure Transducer
Between 0 mmHg and 300 mmHg
±3mmHg
Optional Integrated Temperature*
Temperature
Between 80°F and
110°F (26.7°C and
43.3°C)
All patient populations
±0.2°F
(0.11°C)**
Temperature
Measurement
Site/Mode
Number
of
Subjects
Clinical
Bias (°C)
Limits of
Agreement
Clinical
Repeatability
Oral
106
0.01
0.63
0.14
Rectal
105
-0.12
0.59
0.29