Operator's Manual

Root NIBPT Chapter 14: Troubleshooting
www.masimo.com 200 Masimo
Symptom
Possible Cause
Correction
Home Button does
not work when
pressed.
Internal failure.
Root requires service. See
Return Procedure
on page 247.
Battery does not
charge.
AC power cable
may be
disconnected.
Unplug and replug AC power
cable.
Root Charging
Indicator
illuminates red.
Internal failure.
Root requires service. See
Return Procedure
on page 247.
Nurse Call does not
communicate.
Connection error.
Unplug and replug Nurse Call
connector. See
Nurse Call
Setting Connections
on page
245.
EMR Push
icon
does not appear in
Action Bar.
Wi-Fi is not
enabled.
Root is not
connected to
Patient SafetyNet.
Ensure Wi-Fi is enable. See
Wi-Fi
on page 105. If
EMR Push
icon
does not appear, see
Masimo
Technical Services
on page 248.
Parameter does
not appear in
EMR
Push
screen.
Parameter is not
configured for
EMR Push
on
Patient SafetyNet.
See
Masimo Technical Services
on page 248.