Operator's Manual
www.masimo.com 197 Masimo
Chapter 14: Troubleshooting
Troubleshooting Radical-7, Radius-7, and MOC-9
Modules
For information on troubleshooting values that are provided from Radical-7,
Radius-7, and MOC-9 modules, refer to their respective Instructions for Use
or Operator's Manuals.
Troubleshooting Root
Symptom
Possible Cause
Correction
Root does not turn
on.
Power Button not
pressed long
enough.
Press Power Button for two (2)
seconds.
The battery may
be depleted.
Connect Root to AC power to
charge battery.
One of the fuses is
not operating
properly.
Replace the fuse. See
Replacing
the Fuses
on page 229.