Operator's Manual

Root NIBPT Chapter 13: Messages
www.masimo.com 195 Masimo
Message
Explanation
Next Step
problem persists, replace
probe.
If problem still persists,
contact Customer Service.
Module Error
Module communication
error.
Power cycle the device.
If problem still persists,
contact Customer Service.
Connect Probe
The optional integrated
Temperature Probe has
been disconnected or is not
responsive.
Reconnect Temperature
Probe and try another
measurement cycle.
Check Cuff
(Weak Signal)
Weak or no signal
measured during blood
pressure measurement.
Check that the hose is
connected.
Check that the correct size
cuff is being applied.
Check that the cuff is in the
correct position.
Check that there is no
excessive clothing between
arm and cuff.
Retake another
measurement.
Check Cuff
(Artifact)
Motion may be affecting
ability to take
measurement.
Check Cuff (Out
of Range)
Measurement is out of
range.
Check Cuff
(Measurement
Timeout)
Weak Signal when
measurement is being
taken.