Operator's Manual

Root NIBPT Chapter 13: Messages
www.masimo.com 194 Masimo
Message
Explanation
Next Step
contact Customer Service.
Retake
Measurement
An error has occurred in the
process of taking a patient
temperature measurement
with the optional integrated
temperature module.
Try another measurement
cycle. If problem persists,
replace probe.
Check Ambient
Temperature
Ambient temperature may
be too high or too low.
Verify ambient temperature
does not exceed operating
specification.
Check internal temperature
at probe well.
Replace Probe
Probe is not responsive due
to not being calibrated or
damaged.
Replace probe.
Temperature
Out of Range
Patient or environmental
temperature conditions
may be too low for a
temperature measurement.
Verify patient temperature is
not outside of the
measurement range.
Verify ambient temperature
does not exceed operating
specification.
If conditions are valid and