Operator's Manual

Root NIBPT Chapter 5: NIBP Measurement
www.masimo.com 161 Masimo
*Schedules are configured by authorized personnel.
To measure blood pressure in Schedule mode:
1. Ensure that the correct patient profile is selected before
measurement.
WARNING: Only use Root in Neonatal mode with a neonatal blood
pressure cuff to measure blood pressure on neonates.
2. To change the patient profile, press the Main Menu icon
, then
select Profiles .
3. Press the Profile Name to select the desired patient profile.
4. Properly place the blood pressure cuff on the patient. See
Cuff
Selection and Placement
on page 149.
5. To enable Schedule mode, press the Main Menu icon
, then
select NIBP Settings .
6. On the Intervals screen, change Set Mode to Schedule. The set
mode can also be changed using the action menu.