the diastolic pressure can be determined based upon the rapidly diminishing
amplitude of the pressure fluctuations.
Spot Check Measurement
To spot check measure NIBP
1. Ensure that the correct patient profile is selected before
measurement.
WARNING: Only use Root in Neonatal mode with a neonatal blood
pressure cuff to measure blood pressure on neonates.
Note: The Patient Category determines the initial inflation pressure
of NIBP. Ensure that the proper patient profile and subsequent
patient category are appropriate for the intended patient to be
measured.
2. To change the patient profile, press the Main Menu icon , then
select Profiles .
Press the Profile Name to select the desired patient profile.