possible in the axilla. Do not allow the probe tip to come into
contact with the patient until the probe is placed in the
measurement site. Any prior contact between the probe tip
and the tissue with another material may cause inaccurate
readings. Verify that axillary tissue completely surrounds the
probe tip and place the arm snugly at the patient's side. Firmly
hold the probe in place and keep the tip of the probe in
contact with the tissue throughout the measurement process.
Remove the probe after the temperature measurement is
complete and firmly press the ejection button on the top of
the probe to release the probe cover.
Return the probe to the probe well.
Rectal temperatures are measured using rectal probes which give a
Spot Check Mode (predictive) temperature in approximately 10-13
seconds.
For rectal temperatures, separate the patient's buttocks with
one hand. Use the other hand to gently insert the probe only
5/8 in. (1.5 cm) inside the rectum (less for infants and children).
The use of a lubricant is optional. Tilt the probe so that the tip
is in contact with tissue. Continue to separate the buttocks and
hold the probe in place throughout the measurement process.
Remove the probe after the temperature measurement is
complete and firmly press the ejection button on the top of
the probe to release the probe cover.
Return the probe to the probe well and wash your hands.
Regularly wipe the probe with a cloth dampened with warm water and a
mild detergent solution, a 70% isopropyl alcohol solution, or a 10% chlorine
bleach solution.