Operator's Manual
Root NIBPT Chapter 4: Temperature Measurement
www.masimo.com 144 Masimo
6. Wait for measurement to complete or return the probe to stop
measurement.
Note: If a reliable spot check measurement cannot be made, the
temperature module will automatically switch to continuous mode
to make the measurement.
Note: Once measurement is successfully completed, value is
displayed.
If continuous monitoring is not needed, dock the temperature
probe in the probe well holder to reset measurement.
7. To continuously monitor temperature, touch the action menu
, and then select Continuous mode .
Note: Continuous mode can only be enabled if the probe has not
been removed from the measurement site during spot check.
Note: Continuous mode provides continuous temperature
readings for up to 10 minutes through a direct measurement.
Note: Another spot check cannot be performed until the probe is
first returned to the probe well holder.
8. Dock the temperature probe in the probe well holder to end
temperature monitoring.