Operator's Manual

Root NIBPT Chapter 4: Temperature Measurement
www.masimo.com 143 Masimo
3. Remove the temperature probe from the probe well holder to
initiate spot check.
Note: Ready status will display in the status bar and sound will
play.
4. Apply a single-use disposable probe cover to the temperature
probe before measurement.
WARNING: The use of any other probe cover may produce
temperature measurement errors or may result in inaccurate
readings.
5. Place the temperature probe on the site to begin spot check.
WARNING: Before use, verify the color of the probe cover eject
button on the temperature probe to confirm the proper
application site: Red (rectal), Blue (oral/axillary).
Note: Measuring status displayed in the status bar.