User manual

4
Output cables having a modular plug ( similar to
a telephone connector) must never be connec-
ted to a telephone outlet.
Products with a welded plastic housing are
not repairable. Please contact your supplier for
replacement part.
This product contains hazardous voltages and
there are no user replaceable parts inside the
product. Never attempt to remove the cover.
WARNING: No modification of this equipment is
allowed. Any repair/service should be carried out
by qualified personnel who may get assistance
by contacting the manufacturer or the manu-
facturer’s agent.
Products specified to have automatic polarity
protection must be switched off if a battery is
connected with reverse polarity. The protection
will be automatically reset when the polarity has
been corrected.
In chargers specified to have a replaceable fuse
as polarity protection the fuse must be replaced
if the battery has been connected with reverse
polarity. When replacing the fuse; a fuse of the
same type and rating must be used.
If the product is specified to comply with the
standard for Medical Electrical Equipment (stan-
dards based on IEC60601-1) it complies with
some of the requirements for medical electrical
equipment and may be used in medical applica-
tions and hospital environments.
The product must not be used in the vicinity of
flammable anesthetic gases or in other environ-
ments with flammable or explosive atmosphere.
If the product is specified to comply with the
standard for Medical Electrical Equipment
for Home Healthcare Environment (standard
IEC60601-1-11) it may be used in medical
applications used in a home healthcare environ-
ment.
NOTE: Products relying on safety earth for
protection (Class I) may not be used in home
healthcare environment unless they are perma-
nently wired to the building installation:
Installation must only be carried out by qualified
service personnel, following the below instructions:
- The protective earth conductor must be min.
0.75 mm
2
.
- Connect the protective earth conductor to the
external protective earthing system.
- Verify that the protective earth terminal used is
connected to the external protective earthing
system.
- Verify the integrity of the external protective
earthing system.
This product converts the mains voltage to
a safety extra low voltage.
The output from products applying with 2MOPP
insulation ( model names followed by ”P”) may
be treated as Applied Part Type B or Type BF
according to standard EN/IEC 60601-1 and
may come in physical contact with a patient.The
housing of the product shall not be allowed to
contact the patient.
This product must be operated in an environ-
ment within temperature range +5 to +40°C,
humidity 15 - 93 % RH and atmospheric pres-
sure 70 - 106 kPa (700 - 1060 hPa).
Expected service life of this product and acces-
sories delivered with this product is three (3)
years, if operated as indicated above. However;
the guarantee times indicated in document
”TERMS OF SALES AND DELIVERY FOR MAS-
COT AS” apply (available at www.mascot.com).
Environmental parameters during transport and
storage between uses: temperature range -25 to
+85 °C, humidity 15 - 93% RH NC and atmosp-
heric pressure 70 - 106 kPa (700 - 1060 hPa).
If stored for longer periods of time the
environmental parameters should be within
the temperature range +5 to +35°C, humidity
range 10 - 75% RH NC and atmospheric pres-
sure 70 - 106 kPa (700 - 1060 hPa) to maintain
the products expected service life.
Expected shelf life of this product is one (1) year,
if stored as indicated above.