Instruction Manual

Rotational Vacuum Concentrator RVC 2-33 IR
11 Appendix
Version 08/2009, Rev. 2.1 of 02/12/2014 • sb
89
Translation of the original operating manual
Pos: 217 /20 0 Ch rist/3 71 RV C-BA (PRO JEK TE)/R VC 2- 33 IR /11 0 Anha ng/ 110 -00 40 Ku rzb edien ungs anleit ung RV C 2-3 3 IR @ 31\ mo d_14 074 841 168 65_68.d ocx @ 209 144 @ 2 @ 1
11.4 Brief operating instructions
1. Switch on the system at the mains power switch down on the right side
of the unit.
2. Switch on the cooling trap and/or the vacuum pump.
3. Close the lid.
4. Starting the warm-up phase:
Enter the set values for the warm-up phase into "menu" "options"
"Timer warm-up" und "Temperature warm-up", confirm your entries
and quit the menu.
Press the left-hand function key "mode".
Select the "Warm-up" menu and confirm.
Open the lid when the warm up phase has ended.
Install the rotors and close the lid.
5. Changing the set values for time, temperature, vacuum and safety
pressure:
Select "menu" with the right-hand function key.
Select "Change set values for manual mode" from the main menu.
Select the set value by using the up/down keys and confirm with the
right-hand function key "edit".
Change the set value by using the up/down keys and confirm with
the right-hand function key "ok".
Quit the menu with the left-hand function key.
6. Starting the process "Evaporation manual" or "Evaporation program":
Open the menu "Start with phase..." by pressing the left-hand
function key "mode".
Select "Evaporation manual" or "Evaporation program" using the
up/down keys.
Start the evaporation process by pressing the right-hand function
key.
The function "Evaporation program" can only be selected if a
program has been created before.
The start of the evaporation process starts the following automatic
process run:
- The rotor speed increases (rotating direction left),
- At 70% of the selected maximum speed: the aeration valve closes
automatically (to avoid delays in boiling),
- At 80% of the selected maximum speed: the stop valve opens the
connection to the vacuum pump,
- Further speed increase until the preselected maximum rotor speed
is reached.
7. Interrupting an evaporation process:
Press the left-hand function key "mode".
Select "Stop evaporation" and confirm.
Without an interruption, the evaporation process ends after the
preselected time. The following process runs automatically:
At maximum speed: the stop valve closes the connection to the vacuum
pump, the aeration valve opens; the rotor speed is maintained for 30
sec (aeration phase, to avoid delays in boiling).