Instruction Manual
Rotational Vacuum Concentrator RVC 2-33 IR
8 Maintenance and service
Version 08/2009, Rev. 2.1 of 02/12/2014 • sb
63
Translation of the original operating manual
• Clean the cap and grease the O-ring of the cap slightly with vacuum
grease.
• Install the components or assemblies in a reversed order.
CAUTION
When installing the rotor shaft, ensure that the O-ring used to seal the
rotor chamber is fitted correctly.
• Grease the O-ring with vacuum grease to ensure that it does not get out
of place.
8 O-Ring of the rotor
shaft
Fig. 60: O-ring at the bottom of the rotor shaft
• Plug in the mains plug and switch the rotational vacuum concentrator
on.
• Perform a test run.