Instruction Manual

Rotational Vacuum Concentrator RVC 2-33 IR
8 Maintenance and service
62
Version 08/2009, Rev. 2.1 of 02/12/2014 • sb
Translation of the original operating manual
Remove the rotor shaft incl. magnet coupling.
Fig. 58: Removing the rotor shaft incl. magnet coupling
Remove the four fastening screws on the upper side of the magnetic
coupling using the L-key Torx TX15.
Remove the magnetic coupling from the rotor shaft.
Fig. 59: Removal of the magnet coupling on rotor side
Installation of the rotor shaft and the magnet coupling on rotor side
Clean the rotor chamber, the rotor shaft and the magnet coupling with
cleaning agent or a disinfectant.
CAUTION
Ensure that no cleaning agent or disinfectant gets into the opening at the
bottom of the rotor chamber.
This can lead to irreparable damage at the motor.