8 Maintenance and service
The rotational vacuum concentrator and the accessories are subject to high
mechanical stress. Thorough maintenance performed by the user extends
the service life and prevents premature failure.
CAUTION
If corrosion or other damage occurs due to improper care, the
manufacturer cannot be held liable or subject to any warranty claims.
Use soap water or other water-soluble, mild cleaning agents for
cleaning the RVC and the accessories.
Do not use any corrosive or aggressive substances.
Do not use solvents.
Do not use agents with abrasive particles.
Do not expose the RVC or its accessories to intensive UV radiation (e.g.
sunlight) or thermal stress (e.g. by heat generators).
Pos: 173 /01 0 Uni v ersal mod ule/L eerz ei le @ 0\m od_ 120 211624 450 0_0. doc x @ 11 4 @ @ 1
Pos: 174 /20 0 Ch ris t/3 70 RVC -BA (S TAND AR DMODULE ) /08 0 Wart ung un d Inst and haltu ng/ 080 -001 0 W artu ngsa rbei ten --- ----- --- ----- -- ------ ----- -- ------ -- ----- --- ----- --- ----- -- ------ ----- @ 2 5\m od_1 404 985 235741 _68. doc x @ 18 7527 @ 2 @ 1
8.1 Maintenance
Pos: 175 /20 0 Ch ris t/3 70 RVC -BA (S TAND AR DMODULE ) /08 0 Wart ung un d Inst and haltu ng/ 080 -001 0- 0010 R ot ation s-Vak u um-K onze ntr ator @ 2 5\mod_ 1404 985 236 633 _68. docx @ 18 754 1 @ 3 @ 1
8.1.1 Rotational vacuum concentrator
Switch off the rotational vacuum concentrator via the mains power
switch and disconnect the power cord before cleaning.
Carefully remove all product residues from the rotor chamber using a
cloth.
If the rotational vacuum concentrator has been contaminated with toxic,
radioactive, or pathogenic substances, clean the rotor chamber
immediately with a suitable decontamination agent (depending on the
type of contamination).
WARNING
Take suitable precautions for your own safety if there is a risk of toxic,
radioactive, or pathogenic contamination.
Check the safety guard of the fan (at the front of the bottom side) at
least once a year for dirt and clean it if necessary (e.g. with a vacuum
cleaner).
Open the rotational vacuum concentrator when it is not in use so that
moisture can evaporate.