Instruction Manual

Rotational Vacuum Concentrator RVC 2-33 IR
7 Malfunctions and error correction
56
Version 08/2009, Rev. 2.1 of 02/12/2014 • sb
Translation of the original operating manual
Pos: 167 /20 0 Ch rist/3 70 RV C-BA (S TAND ARD MODULE )/07 0 S törung en u nd Fehl ersuc he/ 070-00 10- 005 0 Kleinfl ans ch ver bindu ngen @ 25\ mod_ 140 498519 019 9_6 8.docx @ 1 874 85 @ 3 @ 1
7.1.5 Small flange connections
Leakages are often due to improper small flange connections between the
various components and hose connections or to leakages in the valves.
Loosen the connection and place the centring ring (with sealing ring
inside) in a centred manner between the flange connections.
Seal the connection with the clamping ring or clamping chain by
tightening the wing nut.
Ensure that the centring ring neither slips out of place nor gets jammed.
Fig. 48: Small flange and centring ring Fig. 49: Small flange with centring ring
and small flange
Fig. 50: Attaching the clamping ring Fig. 51: Tightened clamping ring