Instruction Manual

Rotational Vacuum Concentrator RVC 2-33 IR
6 Operation
Version 08/2009, Rev. 2.1 of 02/12/2014 • sb
49
Translation of the original operating manual
Pos: 149 /20 0 Ch rist/ 370 R VC-B A (S TA NDA RDMODU LE) /06 0 Betri eb/CD pl us/06 0- 0050 -0030 -005 0 O ptio nen (auße r 2- 18) @ 2 5\m od_ 1404 985 153904 _68. docx @ 18 737 3 @ 4 @ 1
6.5.3.5 Options
Fig. 40: Menu “Options”
Change display conrast
Select the menu "Change display contrast".
Change the contrast by pressing the up/down keys.
Confirm the new setting by pressing the right-hand function key.
Fig. 41: Changing the contrast
Select language
The CDplus control system can be used in German or English.
Select the menu "Select language".
Select the language using the up/down keys.
Confirm the input by pressing the right-hand function key.
Fig. 42: Selecting a language
Settings
The Settings menu is used to customise the operation and process
management of the control system.
Select the value you want to set.
Change the value using the up/down keys.
Confirm the input by pressing the right-hand function key.
Fig. 43: Settings menu
Timer Warm-up
You can set a warm-up time between 5 and 99 minutes.