Instruction Manual

Rotational Vacuum Concentrator RVC 2-33 IR
5 Set-up and connection
28
Version 08/2009, Rev. 2.1 of 02/12/2014 • sb
Translation of the original operating manual
3. Connection for aeration and air injection with inert gas
1 Aeration hose connection
2 Air injection hose nozzle
3 Fastener for non-required connection element; here: air filter
CAUTION
If gas (e.g. inert gas) is supplied through the hose nozzle for aeration or for
“air injection”, the threads of the hose nozzles must be sealed with thread
sealant, e.g. Loctite
®
542 (Loctite
®
part number: 54223, bottle with 10ml).
Pos: 95 / 010 Uni ve rsalm odul e/Leerz eil e @ 0\mo d_1 202116 244 500 _0.doc x @ 114 @ @ 1
Pos: 96 / 010 Uni ve rsalm odul e/Leerz eil e @ 0\mo d_1 202116 244 500 _0.doc x @ 114 @ @ 1
Pos: 97 / 200 Chris t /37 0 RVC-B A (S TAND ARD MOD ULE )/050 A u fstell u ng und A ns chluss / 050-00 40 V akuu mve rbi ndung en (au ßer 2- 18 HC L) @ 25\ mod _14 0498 508 7233_6 8.do c x @ 1 87051 @ 2 @ 1
5.4 Vacuum connections
The vacuum connection is realised by way of a vacuum hose with standard
flange connections, clamping rings, or chains, and centring rings.
NOTE
The small flange connections must be installed correctly in order to
prevent leaks (see chapter 7.1.5 - "Small flange connections ").
1 Centring ring
2 Clamping ring
3 Flange connection
Fig. 1: Connecting pieces for vacuum connection