3.7 Safety devices
Pos: 61 / 200 Christ /37 0 RV C-B A (STA NDA RDM ODU LE)/ 030 Si ch erhei t/0 30-007 0-0 010 D eck elver rieg el ung @ 2 5\m od_ 1404 985 013844 _68. docx @ 18 686 9 @ 3 @ 1
3.7.1 Lid lock device
The RVC can only be started when the lid is properly closed. The lid can
only be opened when the rotor has stopped. If the lid is open, the RVC
cannot be started.
Pos: 62 / 010 Uni ve rsalm odul e/Leerz eil e @ 0\mo d_1 202116 244 500 _0.doc x @ 114 @ @ 1
Pos: 63 / 200 Christ /37 0 RV C-B A (STA NDA RDM ODU LE)/ 030 Si ch erhei t/0 30-007 0-0 020 S yste mkon troll e @ 2 5\m od_1 404 985 015 206_68. doc x @ 18 688 3 @ 3 @ 1
3.7.2 System check
An internal system check monitors the data transfer and sensor signals with
regard to plausibility. Errors are detected by continuous self-monitoring of
the system. Error messages are displayed in the process & equipment info
menu (see chapter 6.5.3.4 - "Process and equipment information" or
chapter 7.1.1 - "Error messages").
Pos: 64 / 010 Uni ve rsalm odul e/Leerz eil e @ 0\mo d_1 202116 244 500 _0.doc x @ 114 @ @ 1
Pos: 65 / 200 Christ /37 0 RV C-BA (STAN DAR DM ODULE )/030 S ich erh eit/0 30- 007 0-0030 Netz ausf all siche rheit (au ße r 2-18 ) @ 25\ mod_ 140 498 5016 253_68 .doc x @ 18 689 7 @ 3 @ 1
3.7.3 Power failure safety
The unit stores the current process parameters cyclically. In the event of a
power failure, the electromagnetic stop valve will close automatically, the
rotor chamber will be aerated by the aeration valve and the rotor
decelerates brakeless. After the restart of the power supply, the
evaporation process continues with the last stored parameters.
Pos: 66 / 010 Uni ve rsalm odul e/Leerz eil e @ 0\mo d_1 202116 244 500 _0.doc x @ 114 @ @ 1
Pos: 67 / 200 Christ /37 0 RV C-BA (S TAND ARD MOD ULE)/ 030 Si ch erhei t/0 30-007 0-0 040 S chut zleit erpr üfun g @ 25\ mod_ 140 498 501 7329_6 8.docx @ 1 869 11 @ 3 @ 1
3.7.4 Earth conductor check
For the earth conductor check, there is an equipotential bonding screw on
the rear panel of the RVC. An earth conductor check can be carried out
with the aid of a suitable measuring instrument.
Pos: 68 / 010 U nive rsalm odul e/S eite nwechs el @ 0 \mo d_1 202116 244 312_ 0.d ocx @ 105 @ @ 1