Owner manual

RVC 2-25 CD&
11 Appendix
Version 11/2008, Rev. 2.2 of 04/12/2013 • sb
77
Translation of the original operating manual
Pos: 220 /20 0 Ch rist/3 71 RV C-BA (PROJ EK TE )/RVC 2- 25 CD pl us/1 10 An han g/11 0-0040 Kurz bedi enu ngsa nlei tung @ 8\m od_ 131 5902622 970 _68. docx @ 50 561 @ 2 @ 1
11.4 Brief operating instructions
1. Switch on the system at the mains power switch down on the right side
of the unit.
2. Switch on the cooling trap and/or the vacuum pump.
3. Close the lid.
4. Starting the warm-up phase:
Enter the set values for the warm-up phase into "menu" "options"
"Timer warm-up" und "Temperature warm-up", confirm your entries
and quit the menu.
Press the left-hand function key "mode".
Select the "Warm-up" menu and confirm.
Open the lid when the warm up phase has ended.
Install the rotors and close the lid.
5. Changing the set values for time, temperature, vacuum and safety
pressure:
Select "menu" with the right-hand function key.
Select "Change set values for manual mode" from the main menu.
Select the set value by using the up/down keys and confirm with the
right-hand function key "edit".
Change the set value by using the up/down keys and confirm with
the right-hand function key "ok".
Quit the menu with the left-hand function key.
6. Starting the process "Evaporation manual" or "Evaporation program":
Open the menu "Start with phase..." by pressing the left-hand
function key "mode".
Select "Evaporation manual" or "Evaporation program" using the
up/down keys.
Start the evaporation process by pressing the right-hand function
key.
The function "Evaporation program" can only be selected if a
program has been created before.
The start of the evaporation process starts the following automatic
process run:
The rotor speed increases (rotating direction left)
At 70% of the selected maximum speed: the aeration valve closes
automatically (to avoid delays in boiling)
At 80% of the selected maximum speed: the stop valve opens the
connection to the vacuum pump
Further speed increase until the preselected maximum rotor speed
is reached.