7 Malfunctions and error correction
54
Version 11/2008, Rev. 2.2 of 04/12/2013 • sb
Translation of the original operating manual
Fig. 46: Position of the unit over the edge of the table
Unscrew the two black plastic screws (approx. 8 mm).
Remove the stoppers, e.g. with a screwdriver. The stoppers are
connected to a mechanical release string (see the illustration below).
1 Release string
2 Plastic screw
3 Stopper
Fig. 47: Release string with stopper and screw
Pull the stoppers with their strings vertically downwards until the lid
locks unlock. Unlock the right and left lock in this way.
Fig. 48: Unlocking the lid locks
Open the lid.
After the emergency lid release has been used, reinstall both stoppers
and strings in the reversed order.
Try to locate the error source and contact the service if necessary.