Owner manual

RVC 2-25 CD&
7 Malfunctions and error correction
Version 11/2008, Rev. 2.2 of 04/12/2013 • sb
53
Translation of the original operating manual
Pos: 164 /20 0 C hri st/3 70 RV C-B A (S TA NDA RD MODULE ) /07 0 Stör ung en u nd Fehl ersuc he/ 070 -00 10- 003 0 Syst e mmel du ngen @ 8\ mod_1 309 9391 353 52_ 68.d ocx @ 472 01 @ 3 @ 1
7.1.3 System messages
Error Possible reason Correction
Invalid system type Contact the after-sales service
(see chapter 7.2 - "Service
contact").
IO communication error 0x21 Failure of the CDplus IO module
Switch the system off and then
on again. If the error occurs
again, contact the after-sales
service (see chapter 7.2 -
"Service contact")
IO communication error 0x30 Failure of the reading unit of the
electronic rotor system
Switch the system off and then
on again. If the error occurs
again, contact the after-sales
service (see chapter 7.2 -
"Service contact")
IO communication error 0x40 Failure of the motor control
system
Switch the system off and then
on again. If the error occurs
again, contact the after-sales
service (see chapter 7.2 -
"Service contact")
IO communication error 0x41
Failure of the imbalance
detection module
Switch the system off and then
on again. If the error occurs
again, contact the after-sales
service (see chapter 7.2 -
"Service contact")
Pos: 165 /01 0 Univ ersal module/ Lee rzeile @ 0\ mo d_12021 162445 00_ 0.docx @ 1 14 @ @ 1
Pos: 166 /01 0 Univ ersal module/ Lee rzeile @ 0\ mo d_12021 162445 00_ 0.docx @ 1 14 @ @ 1
Pos: 167 /20 0 Ch rist/3 71 RV C-B A (PR OJEK TE)/R VC 2- 25 CD pl us/0 70 Stö run gen und F ehle rsuch e/07 0-0 010 -0040 Note ntrie gel ung d es Dec k els @ 8\m od_ 1315902 348 228 _68. docx @ 50 533 @ 3 @ 1
7.1.4 Emergency lid release
In the event of a power failure, it is possible to manually open the lid of the
rotational vacuum concentrator.
WARNING
Do not actuate the emergency lid release before the rotor has stopped
and the pressure has been relieved.
Do not use the emergency lid release to open the lid during normal
operation. Unlocked lid locks increase the risk of accidents!
<dg_ref _sou rce_i nli ne >-Pro grammv erw al tun g-
Switch off the mains power switch and disconnect the power cord from
the socket.
Check if the rotor is at a standstill and the rotor chamber is completely
aerated.
Pull the unit a bit over the edge of the table and jack it up (see the
illustration below). Ensure that the centre of gravity of the unit is still on
the table. If necessary, the unit must be secured by a second person.