11.3 Brief operating instructions
1. Switch the system on via the mains power switch on the right side of the
unit.
2. Switch the vacuum pump and/or the cooling trap on.
3. Close the lid.
4. Starting the warm-up phase:
Enter the set values for the warm-up phase under "menu"
"options" "timer warm-up" and "temperature warm-up". Confirm
your entries and quit the menu.
Press the left-hand function key "mode".
Select the "warm-up" menu and confirm.
Open the lid when the warm-up phase has ended.
Install the rotors and close the lid.
5. Changing the set values for time, temperature, vacuum and safety
pressure:
Select "menu" with the right-hand function key.
Select "change set values for manual mode" from the main menu.
Select the set value with the aid of the up/down keys and confirm
with the right-hand function key "edit".
Change the set value with the aid of the up/down keys and confirm
with the right-hand function key "ok".
Quit the menu with the left-hand function key.
6. Starting the process "evaporation manual":
Open the "mode" menu with the left-hand function key.
Select "evaporation manual" using the up/down keys.
Start the evaporation process by pressing the right-hand function
key "enter".
The start of the evaporation process starts the following automatic
process run:
The rotor speed increases (direction of rotation: anti-clockwise)
At 70% of the selected maximum speed: the aeration valve closes
automatically (to avoid boiling retardation).
At 80% of the selected maximum speed: the stop valve opens the
connection to the vacuum pump
Further speed increase until the preselected maximum rotor speed
is reached.
7. Interrupting an evaporation process:
Press the left-hand function key "mode".
Select "stop evaporation" and confirm with the right-hand function
key.
Without an interruption, the evaporation process ends after the
preselected time. The following process runs automatically:
At maximum speed: the stop valve closes the connection to the
vacuum pump, the aeration valve opens; the rotor speed is
maintained for 30 sec (aeration phase, to avoid boiling retardation).
The rotor decelerates until it is at a standstill. After the standstill of
the rotor the lid can be opened in order to remove the rotors.