7 Malfunctions and error correction
48
Version 04/2012, Rev. 1.4 of 25/11/2013 • sb
Translation of the original operating manual
Pos: 146 /20 0 Ch rist/3 71 RV C-BA (PROJ EK TE )/RVC 2- 18 CD pl us_ 2-1 8 CDplus _HC l/07 0 Stö run gen und Fe hlers uche /070 -00 10 -0020 Pr ozes s meld ungen R VC 2 -18 _2-18 H Cl @ 12\ mod_1 3359 398 649 60_ 68.d ocx @ 64149 @ 3 @ 1
7.2.1 Process messages
Error Possible reason Correction
Mains power failure • A mains power failure will be
detected if the power supply is
interrupted under normal
conditions (run mode).
• The process will continue once
the mains power is available
again. The process times will be
reset to zero. The set control
values will be preserved.
Factory settings loaded • All parameters are reset to the
delivery status of the system.
• The factory settings will be
loaded if the corresponding
menu command is executed in
the service menu or if an error
occurred in the parameter
memory.
Nominal rotor speed not reached • After two minutes, the rotor has
still not reached 90% of its
nominal speed.
– set speed is too high
– drive moves sluggishly
• Check the set speed.
• When the rotor is at a standstill
and the system is switched off,
check whether the drive moves
sluggishly, by hand. Contact the
after-sales service (see chapter
7.3 - "Service contact").
Safety pressure not reached • The safety pressure inside the
rotor chamber must be reached
within 5 minutes after the start
of an evaporation process.
• Check whether the value that
was entered for the safety
pressure is not too low.
• Check the components for any
possible leaks.
Pos: 147 /01 0 Univ ersal module/ Lee rzeile @ 0\ mo d_12021 162445 00_ 0.docx @ 1 14 @ @ 1
Pos: 148 /20 0 C hri st/3 70 RV C-B A (S TA NDA RD MODULE ) /07 0 Stör ung en u nd Fehl ersuc he/ 070 -00 10- 003 0 Syst e mmel du ngen @ 8\ mod_1 309 9391 353 52_ 68.d ocx @ 472 01 @ 3 @ 1
7.2.2 System messages
Error Possible reason Correction
Invalid system type • Contact the after-sales service
(see chapter 7.3 - "Service
contact").
IO communication error 0x21 • Failure of the CDplus IO module
• Switch the system off and then
on again. If the error occurs
again, contact the after-sales
service (see chapter 7.3 -
"Service contact")
IO communication error 0x30 • Failure of the reading unit of the
electronic rotor system
• Switch the system off and then
on again. If the error occurs
again, contact the after-sales
service (see chapter 7.3 -
"Service contact")
IO communication error 0x40 • Failure of the motor control
system
• Switch the system off and then
on again. If the error occurs
again, contact the after-sales
service (see chapter 7.3 -
"Service contact")
IO communication error 0x41
• Failure of the imbalance
detection module
• Switch the system off and then
on again. If the error occurs
again, contact the after-sales
service (see chapter 7.3 -
"Service contact")