7 Malfunctions and error correction
Version 04/2012, Rev. 1.4 of 25/11/2013 • sb
47
Translation of the original operating manual
Pos: 141 /20 0 Chri st/3 70 RVC-B A (S TAN DAR DMODULE ) /070 S törung en u nd Fehl ersuc he/ 070 S törung en u nd Fehl ersuc he === ==== ===== ===== == ===== === @ 8\mod_ 1309 939 131 742_68. docx @ 47 129 @ 1 @ 1
7 Malfunctions and error correction
Pos: 142 /20 0 Ch rist/3 70 RVC -BA (S TA NDA RDMOD ULE) /07 0 Stör ung en und Fehl ersuc he/ 070 -0010 All g emei ne Stö rungen @ 8\ mod _13099 413 440 43_ 68.doc x @ 472 24 @ 2 @ 1
7.1 General malfunctions
Malfunctions are displayed in the menu "Process & equipment info" (see
chapter 6.5.3.2 - "Process and equipment information"). A sound signal
sounds when an error message is generated.
Eliminate the source of the problem (see the following chapters).
Acknowledge the error messages by pressing the right-hand function
key.
Pos: 143 /01 0 Univ ersal module/ Lee rzeile @ 0\ mo d_12021 162445 00_ 0.docx @ 1 14 @ @ 1
Pos: 144 /20 0 Chri st/3 71 RVC-B A (PR OJEK TE)/R VC 2- 18 CD plus_ 2-1 8 CDpl us_H Cl/07 0 Stö run gen und Fe hlers uche /070 -0010 -001 0 Fehl er meldu ngen RV C 2- 18_2-1 8 HCl @ 12\ mod _13 3593986 448 7_6 8.do cx @ 6414 1 @ 3 @ 1
7.2 Error messages
Error Possible reason Correction
Lid cannot be opened The rotor chamber is not
aerated completely
Aerate the rotor chamber
completely
Lid contacts error The system has detected an
open lid during the evaporation.
The rotor will be stopped
immediately. The lid must be
closed while the rotor is moving.
Check the lid contacts. If
necessary, contact the after-
sales service (see chapter 7.3 -
"Service contact")
Rotor overtemperature The maximum permissible rotor
temperature has been
exceeded
If the system does not cool
down, disconnect it immediately
from the power supply.
Contact the after-sales service
(see chapter 7.3 - "Service
contact")
Chamber overtemperature The maximum permissible
chamber temperature has been
exceeded.
If the system does not cool
down, disconnect it immediately
from the power supply.
Contact the after-sales-service
(see chapter 7.3 - "Service
contact").
Permissible imbalance exceeded The rotor will be stopped
immediately, the process will
also be stopped.
improper loading of the
rotor
glass breakage during the
evaporation
uneven evaporation
Eliminate the imbalance and
restart the unit. If the error is still
present, contact the after sales-
service (see chapter 7.3 -
"Service contact")
VSK3000 calibration error Incorrect calibration of the
vacuum sensor VSK3000
The sensor must be recalibrated
(see the separate operating
manual).