8.1.2 Rotor chamber
If the rotor chamber is contaminated or if glass is broken, the rotor
chamber, the rotor shaft and the rotor must be cleaned immediately in order
to prevent corrosion and damage to the ball bearings.
Cleaning of the rotor chamber
1 Rotor chamber
2 Cap
3 Rotor shaft
Fig. 41: Rotor chamber with rotor shaft
• Open the lid.
• Disconnect the power cord from the wall outlet or instrument receptacle
before cleaning.
DANGER
Hot surface!
Let the rotor chamber cool down to room temperature!
• Pull off the cap.
• Remove the countersunk screw with the supplied hexagon socket key
(size 2.5).