5 Set-up and connection
Pos: 82 / 200 Chris t/ 370 R VC-B A (S TANDA RDM OD ULE)/ 050 A uf stellu ng u nd Ans c hluss/ 050 -00 10 A ufstel lo rt, Ei ns atzo rt @ 8\mod_ 130 9335 833 534 _68 .docx @ 46 605 @ 2 @ 1
5.1 Installation site
Operate the RVC only in closed and dry rooms.
• The table must be stable and have a solid, even tabletop.
• Ensure sufficient ventilation. Do not place any paper, cloth, or similar
material behind or under the unit, since otherwise the air circulation will
be impaired.
• Keep a safety distance of at least 30 cm from the wall so that the vents
in the unit remain fully effective.
• The ambient temperature must be in the range of +10°C to +25°C.
• Do not subject the RVC to thermal stress, e.g. by positioning it near
heat generators.
• Avoid direct sunlight (UV radiation).
Pos: 83 /010 Unive rsalm odule/ L ee rzeile @ 0\ mod _120211 624450 0_0.docx @ 11 4 @ @ 1
Pos: 84 / 200 Christ / 370 R VC-B A (STA NDA RDM OD ULE)/ 050 Au fs tellu ng u nd Ans c hluss/ 050 -00 20 E ner gieve rsor gun g---- --- ----- ----- --- ----- -- ------ ----- - @ 8\mod_ 1309 335 833 862 _68. docx @ 46 613 @ 2 @ 1
5.2 Power supply
Pos: 85 / 200 Chris t /37 1 RVC -BA (P ROJE KTE )/R VC 2-1 8 CD pl us_2 -18 CD pl us_HCl /050 Aufs tell ung und Ansc hluss /05 0-0 020 -00 10 An schlus s art RVC 2-1 8 @ 12\m od_ 133 697 507 2135_6 8.doc x @ 6 4455 @ 3 @ 1
5.2.1 Connection
DANGER
The operating voltage on the name plate must correspond to the local
supply voltage
christ rotational vacuum concentrators are units of safety class
I.Units of the type RVC 2-18 CD
plus
have a three-wire power cord with an
IEC C13 connector (see chapter 10 - "Technical data").
An equipotential bonding screw is located on the back (see chapter 2.1.1 -
"Functional and operating elements"). This equipotential bonding screw
can be used to perform an earth conductor check.
Pos: 86 /010 Unive rsalm odule/ L ee rzeile @ 0\ mod _120211 624450 0_0.docx @ 11 4 @ @ 1
Pos: 87 / 200 Christ /37 0 RV C-BA (STAN DAR DM ODULE )/ 050 Au fstell ung u nd Ans c hluss/ 050 -00 20- 0020 Si ch erunge n ba useits @ 8\ mod _13 093 3583 4596_6 8.do cx @ 4662 9 @ 3 @ 1
5.2.2 Customer-provided fuses
Typically, the RVC must be protected with 16 Amp G fuses that are to be
provided by the customer.