8.1.1.1 Ice condenser chamber
Before each start-up, ensure that the ice condenser chamber is free from
water residues.
• Open the media drain valve to drain off any liquid. Then, close the
valve.
• If necessary, wipe the ice condenser chamber dry with a cloth.
Pos: 201 /01 0 Uni v ersal mod ule/L eerz ei le @ 0\m od_ 120 211624 450 0_0. doc x @ 11 4 @ @ 1
Pos: 202 /20 0 Ch rist/3 61 G T-BA L abo r-Pil ot (P ROJE KTE)/E psil on S ta ndard/0 80 Wa rtu ng u nd Ins tandhal tung /08 0-0 010-00 10- 002 1 Sonder auss tat tung: Tüv erschl us srie gel aus Edels tahl E psilon @ 25 \mo d_1404 9846 771 36_ 68.d ocx @ 185 889 @ 4 @ 1
8.1.1.2 Special equipment: Stainless steel door latch at the loading door
Systems that are equipped with a stainless-steel or acrylic glass door with a
stainless-steel door latch have a slide pad on the door latch. This slide pad
is included in the scope of supply as a spare part. It must be inspected
regularly for signs of wear. When its thickness is < 0.5 mm, it must be
replaced.
1 Door latch
2 Slide pad
(part no. 177270)
3 Torx screws
Fig. 80: Door latch and slide pad
Replacing the slide pad:
Unscrew the two screws of the slide pad with a Torx
®
TX10 screwdriver,
remove the old slide pad, and install the new one.