Creating a blank program (1)
This button is used to call up a blank program template. Only section 1 is
pre-defined as the loading section. In this phase, the start conditions of the
programs are defined. The room temperature (20°C) is the standard
default.
NOTE
If the product is frozen outside the freeze-dryer (double-chamber method),
the shelf temperature must be adapted accordingly in section 1.
For all the other sections, the set values must be defined (see figure
below):
Press the button “Insert section” (4) and select the position of the
section as well as the freeze-drying phase. The section will be inserted
at the defined position.
Adapt the parameters of the inserted section. All of the values are not
available in all of the freeze-drying phases.
Insert and edit the next section.
NOTE
During the creation of a program, the order of the individual freeze-drying
phases must be maintained. This means that it is not possible to insert a
freezing section after a main drying section.
“Program” buttons:
1 "Program"
2 "Program name"
3 "Show diagram"
4 "Insert section"
5 "Delete section"
6 Button for calling up
more functions and
set values
7 "Save"
Fig. 67: Editing a blank program template
The button “Delete section” (5) can be used to delete sections.
The button ">>more" (6) can be used to enter more functions and set
values in a program-related manner, depending on the type of freeze-
dryer (see the sections below).
Proceed in this manner in order to create an entire program according
to your specific needs.
Pressing the button “Diagram(3) displays the program in the form of a
diagram.