6 Operation
Version 04/2013, Rev. 1.5 of 16/12/2014 • sb
59
Translation of the original operating manual
Fig. 44: Dialog box "Network"
System properties (3)
This dialog box is used to change various system settings.
Fig. 45: Dialog box "System properties"
Beeper: The beeper sounds in the event of a malfunction, for example.
If the setting is "On", the beeper sounds every few seconds until the
user acknowledges the message.
If the setting is "Silent", the beeper sounds once when the malfunction
occurs.
If the setting is "Off", the beeper will not sound at all.
Click on touch: If this function is active, a clicking sound can be heard
whenever the system registers that a button has been touched.
Behaviour of the pressure control valve in the case of insufficient ice
condenser cooling: If this function is active (button "closed"), the pressure
control valve will close at an ice condenser temperature of
-20°C during the drying process in order to avoid damage to the vacuum
pump caused by the withdrawal of condensable gases. A corresponding
error message will be displayed. Pressing the button again deactivates the
function (button "controlled").