5 Set-up and connection
Version 04/2013, Rev. 1.2 of 20/02/2014 • sb
35
Translation of the original operating manual
1 Vacuum sensor
2 Connection socket
Fig. 15: Position of the vacuum sensor and the connection socket
Switch the unit off by actuating the mains power switch.
Take the vacuum sensor out of its original packaging and fasten it to the
connector with a clamping ring (DIN16KF) and a centring ring (included
in the scope of supply).
Plug the connector to the connection socket and hand-tighten the
screws on the connector.
NOTE
It is absolutely essential to comply with the manufacturer’s instructions in
the separate operating manual of the vacuum sensor!
Thyracont VSP62/63
MKS 722B
Pfeiffer TPR 280/281
Fig. 16: Installed vacuum sensors of different manufacturers