Operating Manual Freeze-dryer Delta 1-24 LSC& Part no. 102300, 102302, 102304, 102306, 102308, 102310, 102316, 102318, 102320, 102322 Delta 2-24 LSC& Part no. 102301, 102303, 102305, 102307, 102309, 102311, 102317, 102319, 102321, 102323 Translation of the original operating manual sb Version 04/2013 Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 2 Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& In case of inquiries, please state the following numbers: Order number Serial number © Copyright by Martin Christ Gefriertrocknungsanlagen GmbH An der Unteren Söse 50 37520 Osterode am Harz Germany Tel.: +49 (0) 5522 / 5007-0 Fax: +49 (0) 5522 / 5007-12 Web: www.martinchrist.de E-mail: info@martinchrist.de 4 Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& Table of contents 1 General information ........................................................................................................... 10 1.1 Importance of the operating manual................................................................................ 10 1.2 Intended use ................................................................................................................... 10 1.3 Warranty and liability......................................
Delta 1-24 LSC& Delta 2-24 LSC& Table of contents 4 Storage and transport ....................................................................................................... 30 4.1 Dimensions and weight ................................................................................................... 30 4.2 Storage conditions .......................................................................................................... 30 4.3 Notes on transport .......................................
Delta 1-24 LSC& Delta 2-24 LSC& Table of contents 7 Malfunctions and error correction .................................................................................... 77 7.1 General malfunctions ...................................................................................................... 77 7.1.1 Power failure.............................................................................................................. 78 7.1.2 Insufficient vacuum ..............................................
Delta 1-24 LSC& Delta 2-24 LSC& 1 General information Pos: 1 /2 00 C hrist/ 360 GT -BA Lab or-Pilot (STANDARDM ODULE)/ 010 Allg emei ne I nfor mati one n/01 0 Allge mein e In for matio nen == == === == == === == == == === == == === == @ 5\ mod _12 991 361 501 10_6 8.d ocx @ 344 91 @ 1 @ 1 1 General information Pos: 2 /2 00 C hrist/ 360 GT -BA Lab or-Pilot (STANDARDM ODULE)/ 010 Allg emei ne I nfor mati one n/01 0-0 010 Stelle nwert de r Betri ebsa nleitu ng @ 5\m od_ 129 913 615 061 0_68 .
Delta 1-24 LSC& Delta 2-24 LSC& 1 General information Pos: 6 /2 00 C hrist/ 360 GT -BA Lab or-Pilot (STANDARDM ODULE)/ 010 Allg emei ne I nfor mati one n/01 0-0 030 Gew ährl eistun g u nd Ha ftun g @ 5 \mo d_1 299 136 151 485 _68. docx @ 34 515 @ 2 @ 1 1.3 Warranty and liability The warranty and liability are subject to our "General Terms and Conditions" that were distributed to the operator upon the conclusion of the contract.
Delta 1-24 LSC& Delta 2-24 LSC& 2 Layout and mode of operation Pos: 17 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /02 0 Aufb au u nd Wirk ungs weise/ 020 Aufb au u nd Wir kung sweise == === == == === == == === == == === = @ 5\m od_ 129 913 628 8294 _68 .
Delta 1-24 LSC& Delta 2-24 LSC& 2 Layout and mode of operation 5 Contact bolt 6 Pipe connection of the vacuum pump (behind the cover plate) Ice condenser Vacuum sensor 7 8 Fig. 3: Ice condenser chamber 9 Control system on/off (mains power switch) 10 Steerable castors Fig. 4: Right side of the freeze-dryer Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 2 Layout and mode of operation 11 Aeration valve 12 Media drain valve 13 Fixed castors Fig. 5: Left side of the freeze-dryer 14 Vacuum sensor 15 Mains voltage (Main switch) 16 Electrical connection of the vacuum sensor 17 Serial interface 18 Mains cable 19 Heat exchanger of the refrigeration unit (behind the panel) 20 Option: Connection for electrical lifting device 21 Name plate (see chapter 2.1.2 - "Name plate") 22 Vacuum connection Fig.
Delta 1-24 LSC& Delta 2-24 LSC& 2 Layout and mode of operation Pos: 21 / 200 Christ /36 1 GT -BA La bor -Pilot (PROJEKTE )/Delt a 1- 24_ 2-2 4 LSCpl us/02 0 Aufb au und Wi rkun gsweise/ 020 -00 10- 002 0 Type nschild Delta @ 1 4\mo d_1 354 865 314 837 _68. docx @ 76 720 @ 3 @ 1 2.1.
Delta 1-24 LSC& Delta 2-24 LSC& 2 Layout and mode of operation Pos: 23 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /02 0 Aufb au u nd Wirk ungs weise/ 020 -00 20 Wi rkun gsweise -- --- --- -- --- -- --- -- --- - @ 5\mo d_1 299 136 292 466 _68. docx @ 34 587 @ 2 @ 1 2.2 Mode of operation Pos: 24 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /02 0 Aufb au u nd Wirk ungs weise/ 020 -00 20- 001 0 G efrie rtr ocknu ng allgem ein @ 5\ mod_ 129 913 629 284 0_6 8.
Delta 1-24 LSC& Delta 2-24 LSC& 2 Layout and mode of operation Melting point pressure curve Pressure (mbar) SOLID A Melting B Sublimation 0.37 mbar V Su apo bl ur im p at res io s n ur pr e es cu su rv re e f cu or rv ice e = 6.11 mbar LIQUID ou Vap u ess r pr e urv re c Evaporation Triple point GASEOUS -30°C Temperature (°C) Fig.
Delta 1-24 LSC& Delta 2-24 LSC& 2 Layout and mode of operation Pressure (mbar) 1 bar LIQUID Melting point pressure curve 1 SOLID cur ve Triple point Va p su ou bl r p im re at ss io ur n e pr cu es rv su e re for cu ic rv e = e 2 ou Vap ur e ess r pr GASEOUS 20°C Temperature (°C) Fig. 9: Freeze-drying phases Please find further information about basic principles, optimum procedures and applications in the brochure "Smart freeze-drying", which can be downloaded at www.martinchrist.
Delta 1-24 LSC& Delta 2-24 LSC& 2 Layout and mode of operation Pos: 26 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /02 0 Aufb au u nd Wirk ungs weise/ 020 -00 20- 002 0 Abla uf ei nes G efrie rtr ockn ungs proz esses @ 5\ mod _12 991 362 9329 4_6 8.d ocx @ 346 03 @ 3 @ 1 2.2.
Delta 1-24 LSC& Delta 2-24 LSC& 2 Layout and mode of operation Fig. 10: Shell freezing Fig. 11: Cooling bath with spin freezing device NOTE Pos: 31 / 010 Unive rsalm odul e/ Fig. 12: Spin freezing If the product that is to be dried contains solvents or high salt concentrations, it may start to thaw during the drying process, which is indicated by clearly visible foaming. In order to prevent this, the product must be frozen as deeply as possible, e.g.
Delta 1-24 LSC& Delta 2-24 LSC& 2 Layout and mode of operation The end of the main drying phase is reached, when the product temperature and the shelf temperature are nearly identical. The temperature difference between the shelf and the product should be approximately 3 K to 5 K. Pos: 33 / 010 Unive rsalm odul e/ L ee rzeile @ 0\ mod _120 211 624 450 0_0.
Delta 1-24 LSC& Delta 2-24 LSC& 2 Layout and mode of operation Special equipment: defrosting with hot water For hot-water defrosting, the freeze-dryer must be connected to a hot water supply. NOTE The ice condenser is defrosted by feeding hot water into the ice condenser chamber. The hot water is sprayed onto the ice condenser until the maximum filling level is reached in the ice condenser chamber.
Delta 1-24 LSC& Delta 2-24 LSC& 3 Safety Pos: 43 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /03 0 Sicher heit/ 030 Siche rheit == === == == === == == == === == == === == == === @ 5\ mo d_1 2991 364 608 07_ 68.d ocx @ 346 60 @ 1 @ 1 3 Safety Pos: 44 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /03 0 Sicher heit/ 030 -00 10 B eschilde ru ng d es Ge räts @ 5\ mod _12 991 364 611 97_ 68.d ocx @ 346 68 @ 2 @ 1 3.
Delta 1-24 LSC& Delta 2-24 LSC& 3 Safety Pos: 46 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /03 0 Sicher heit/ 030 -00 20 Sy mbol - un d Hinw eiserkl äru nge n @ 5 \mo d_1 299 136 4616 19_ 68. docx @ 34 676 @ 2 @ 1 3.2 Explanation of the symbols and notes This operating manual uses the following names and symbols to indicate hazards: This symbol stands for a direct hazard to the life and health of persons.
Delta 1-24 LSC& Delta 2-24 LSC& 3 Safety Pos: 48 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /03 0 Sicher heit/ 030 -00 30 V era ntwor tun g des Betr eiber s @ 5 \mo d_1 299 136 4620 25_ 68. docx @ 34 684 @ 2 @ 1 3.3 Responsibility of the operator The operator is responsible to authorise only qualified personnel to work on the freeze-dryer (see chapter 3.4 - "Operating personnel").
Delta 1-24 LSC& Delta 2-24 LSC& 3 Safety Pos: 54 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /03 0 Sicher heit/ 030 -00 60 Sic her heitshi nweise ++ +++ ++ ++ +++ ++ ++ +++ ++ ++ ++ +++ + @ 5\m od_ 129 9136 463 260 _68. docx @ 34 708 @ 2 @ 1 3.6 Safety instructions The following instructions must be observed prior to every drying process: • • WARNING • • • • Pos: 55 / 010 Unive rsalm odul e/ L ee rzeile @ 0\ mod _120 211 624 450 0_0.
Delta 1-24 LSC& Delta 2-24 LSC& 3 Safety Pos: 59 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /03 0 Sicher heit/ 030 -00 60- 002 0 Mech anisch e Siche rheit @ 5\ mod _12 991 364 640 88_ 68.d ocx @ 347 24 @ 3 @ 1 3.6.2 Mechanical safety In order to ensure the safe operation of the freeze-dryer, please comply with the following points: • • • WARNING • • • • • • Pos: 60 / 010 Unive rsalm odul e/ Do not use the freeze-dryer if it was installed incorrectly.
Delta 1-24 LSC& Delta 2-24 LSC& 3 Safety Pos: 65 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /03 0 Sicher heit/ 030 -00 60- 005 0 Ch emisc he u nd biologisc he Sich erh eit @ 5\m od_1 299 136 465 291 _68. docx @ 34 748 @ 3 @ 1 3.6.
Delta 1-24 LSC& Delta 2-24 LSC& 3 Safety Pos: 71 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /03 0 Sicher heit/ 030 -00 81 V erh alten bei G efah re n un d Unf ällen D elta @ 17\ mod _13 655 005 7587 7_6 8.d ocx @ 844 76 @ 2 @ 1 3.
Delta 1-24 LSC& Delta 2-24 LSC& 4 Storage and transport Pos: 76 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /04 0 La ger ung und Tr ans port /040 L age run g un d T rans por t= === == == === == == === == == === == @ 6\ mod _12 991 369 224 11_ 68.
Delta 1-24 LSC& Delta 2-24 LSC& 4 Storage and transport Pos: 83 / 200 Christ /36 1 GT -BA La bor -Pilot (PROJEKTE )/Delt a 1- 24_ 2-2 4 LSCpl us/04 0 La ge rung un d T rans por t/04 0-0 040 Ver packu ng D elta @ 15\ mod _13 553 944 7207 8_6 8.d ocx @ 767 85 @ 2 @ 1 4.4 Packaging The freeze-dryer is packaged in a wooden crate. • After opening the packaging, take out the accessories. • Remove the packaging material. • Remove the side walls of the crate.
Delta 1-24 LSC& Delta 2-24 LSC& 4 Storage and transport Pos: 87 / 200 Christ /36 1 GT -BA La bor -Pilot (PROJEKTE )/Delt a 1- 24_ 2-2 4 LSCpl us/04 0 La ge rung un d T rans por t/04 0-0 060 In ner betri eblich er Tra nspo rt Del ta @ 15\ mod _13 553 9447 264 3_6 8.do cx @ 7 679 3 @ 2 @ 1 4.6 On-site transport The freeze-dryer comes with castors. It can be pushed by a suitable number of persons over short distances. Over longer distances, it should be transported by way of suitable lifting equipment (e.g.
Delta 1-24 LSC& Delta 2-24 LSC& 5 Set-up and connection Pos: 90 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /05 0 Aufstell ung und Anschlus s/05 0 Aufs tellun g u nd Ansc hluss= == == === == == === == == == === == @ 6\ mod _12 991 371 3779 7_6 8.d ocx @ 349 66 @ 1 @ 1 5 Set-up and connection Pos: 91 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /05 0 Aufstell ung und Anschlus s/05 0-0 010 Aufst ellort (Alpha -Delt a) @ 6\m od_ 129 913 709 511 0_6 8.docx @ 3 489 4 @ 2 @ 1 5.
Delta 1-24 LSC& Delta 2-24 LSC& 5 Set-up and connection Pos: 98 / 200 Christ /36 1 GT -BA La bor -Pilot (PROJEKTE )/Delt a 1- 24_ 2-2 4 LSCpl us/05 0 Aufst ellung un d Anschl uss/0 50- 0030 Belüft ungs ventil D elta @ 15\ mo d_13 553 950 309 15_ 68.d ocx @ 780 60 @ 2 @ 1 5.3 Aeration valve The aeration valve is located on top of the left side of the unit (see chapter 2.1.1 - "Functional and operating elements"). After the end of a freeze-drying process, the unit will be aerated via the aeration valve.
Delta 1-24 LSC& Delta 2-24 LSC& 5 Set-up and connection 1 2 Vacuum sensor Connection socket Fig. 15: Position of the vacuum sensor and the connection socket • • • Switch the unit off by actuating the mains power switch. Take the vacuum sensor out of its original packaging and fasten it to the connector with a clamping ring (DIN16KF) and a centring ring (included in the scope of supply). Plug the connector to the connection socket and hand-tighten the screws on the connector.
Delta 1-24 LSC& Delta 2-24 LSC& 5 Set-up and connection The vacuum sensor comes supplied in a calibrated state. NOTE After the freeze-dryer has been switched on, the vacuum sensor needs several minutes until it is ready for operation. Pos: 104 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.
Delta 1-24 LSC& Delta 2-24 LSC& 5 Set-up and connection Pos: 109 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 50 Aufst ellun g un d Anschl uss/0 50- 010 0 G um miventil e @ 9 \mo d_1 319 531 8057 44_ 68. docx @ 52 539 @ 2 @ 1 5.8 Rubber valves The rubber valves (part no. 121860) enable the connection of round-bottom flasks, wide-neck filter bottles, or distributors for ampoules to a manifold or drying chamber. Depending on the connector of the components, the blue plug can be removed.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Pos: 111 /01 0 Univ ersal mod ule/ Sei tenwe chsel @ 0\ mod _12 021 1624 431 2_0 .docx @ 1 05 @ @ 1 Pos: 112 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 60 Bet rieb/ 060 B etrie b= === == == === == == === == == === == == == === == == == @ 6\m od_ 129 913 739 277 9_68 .
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Pos: 121 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 60 Bet rieb/ LSCplus/ 060 -00 50 LSCplus Anl age nsteu eru ng- -- --- -- --- --- -- --- -- --- -- --- --- -- --- -- --- --- @ 6\ mo d_1 2991 425 822 54_ 68.d ocx @ 366 39 @ 2 @ 1 6.5 LSCplus control system The control system LSCplus (Lyo Screen Control plus) was specifically developed for the control of freeze-drying processes.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation The user interface is divided into four main windows that can be called up by touching the corresponding buttons: Manual This window is also the standard user interface. It is used to control the freeze-drying process manually. Program This area is used to create and execute programs for automating the freeze-drying process and for making it reproducible. This function is only available in combination with the programmer module PGMplus option.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Status line (1) This line shows the operating status of the freeze-dryer as well as the active phase. The status line also shows the current date and time. The clock is batterybuffered and must be reset after a failure (chapter 6.5.1.3 - "Main window "Options "", section ”Administration”"). In addition, the drive symbol provides information concerning the status of the external data storage device or of the network drive. The following symbols are possible: ...
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Actual values of the current process (4) This area shows the current process data. The fields can be configured as desired: • Select the button of the field that you would like to reconfigure. A dialog box opens: Fig. 22: Dialog box "Select actual value" • Select the desired configuration or quit the dialog box by pressing the "return" button. In this way, it is possible to configure a personalised overview of the actual values.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Option: USB process recording (see chapter 6.6 - "Optional extensions") • Select the "Process recording" function in the dialog box "Tools". • Select the input fields ("Batch data"). A keyboard for the data input will be displayed. • If necessary, select the “Options” tab, choose between manual or automatic recording, and define a recording interval. • Press the "return" button in order to close the dialog box.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Fig. 25: Dialog box “Pressure increase test” • The “Test stop” button stops the pressure increase test. Dialog box "Schematic system diagram" (6) Pressing the button "Schematic system diagram" displays a schematic diagram of the system on the left-hand side of the screen, including all the components. Active components are displayed in green. Touching a component calls up its name and reference designator. Fig.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Malfunctions are divided into three categories: • Red: error messages • Orange: process messages • Yellow: general messages The representation of the message provides information on its current status. A double frame around a message means that the error has not been eliminated yet. The colour of the button "quit" changes from blue to grey once the message has been acknowledged. Fig.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Details Touching the message calls up details concerning the error message: • Cause of the message • Effects of the message • Measure to eliminate the error • Reference designator Use the arrow keys to call up the individual windows. The error message text is always followed by an error code. Always indicate the error code in the event of enquiries or service requests! NOTE 4 5 6 7 Error message Error code Detailed information Arrow keys Fig.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation "Dialog box "Operating mode: select/start (8) After the set values have been entered for the process, the process can be started manually with this function (see chapter 6.5.2 - "Manual freezedrying"). Fig. 30: Dialog box "Operating mode: select/start" (The version of the dialog box that is displayed depends on the equipment of the freeze-dryer.) Button "Stop" (9) Pressing this button stops the current process. The system switches to the standby status.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation 6.5.1.2 Main window "Program" In the main window "Program", pre-programmed freeze-drying processes can be loaded and edited and new programs can be created with the PGMplus programmer module. For this function, the PGMplus programmer module must be available. NOTE The PGMplus programmer module is an option that is not included as standard (see chapter 6.5.3 - "Option: freeze-drying with the PGMplus programmer module").
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Pos: 130 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 60 Bet rieb/ LSCplus/ 060 -00 50- 001 0-0 030 Hau ptfe nster Opti one n La bor @ 6\ mod _12 991 427 521 08_ 68.d ocx @ 366 79 @ 4 @ 1 6.5.1.3 Main window "Options " The main window "Options" is used to perform fundamental basic settings of the control system in order to adapt it perfectly to the respective area of application of the freeze-dryer.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Change scale unit (2) This dialog box is used to change the unit of measurement for the temperature and vacuum. Fig. 34: Dialog box "Change scale unit" Freeze-drying process (3) This function depends on the unit type and is not available for all types of freeze-dryers. NOTE Prior to the start of the process, the correct process must be selected.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Administration “Administration” options: 1 "Date/Time" 2 "Network" 3 "System properties" 4 "Access protection" 5 "Device options" 6 "Configuration of sensors" Abb. 36: Dialogfenster "Optionen" / Verwaltung Date/Time (1) The LSCplus control system is equipped with an integrated, battery-buffered clock. After a failure of the buffer battery, the date and time must be reset. Fig. 37: Dialog box “Change date/time Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Network (2) This dialog box is used to change various system settings. The modifications will not become effective until after a restart of the unit. NOTE Fig. 38: Dialog box "Network" System properties (3) This dialog box is used to change various system settings. Fig. 39: Dialog box "System properties" Beeper: The beeper sounds in the event of a malfunction, for example.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Click on touch: If this function is active, a clicking sound can be heard whenever the system registers that a button has been touched. Behaviour of the pressure control valve in the case of insufficient ice condenser cooling: If this function is active (button "closed"), the pressure control valve will close at an ice condenser temperature of 20°C during the drying process in order to avoid damage to the vacuum pump caused by the withdrawal of condensable gases.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation • The button with the lock symbol blocks the access immediately and the system switches to the default setting. Fig. 42: Access blocked, the data cannot be edited Operator/maintenance/administrator password: For each of these access levels, certain editing rights have been defined. They can be enabled with the corresponding password. The rights of the various access levels are detailed in the following table.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation If the freeze-dryer is to be extended by an option, this option must be enabled via this dialog box. • Press the button "Add option". An input window opens. • Enter the six-digit Christ activation code that was supplied for this option. Note that the keys are case sensitive. Options can be removed in the same way. The modifications will not become effective until after a restart of the unit.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Service “Service” options: 1 "Operating hours" 2 "Failure memory" 3 "Wireless shelf technology" 4 "Vacuum pump oil change" "Maintenance" "Sensor adjustment" 5 6 Fig. 45: Dialog box "Service" (varies depending on the type of system) Operating hours (1) This dialog box is used to call up the number of operating hours of the various components of the freeze-dryer, e.g. the refrigeration unit, vacuum pump, or pressure control valve.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Always indicate the error code in the event of enquiries or service requests! NOTE Fig. 47: Dialog box "Failure memory" Wireless shelf technology (3) This function depends on the unit type and is not available for all types of freeze-dryers. NOTE The wireless shelf technology enables the wireless control and monitoring of the shelf temperature. For this purpose, an address must be assigned to the WST modules of the shelves.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation • • • In order to assign an address, open the dialog box "Wireless Shelf Technology". Ensure that only the WST module to which an address is to be assigned is connected. If several modules are connected, they will all receive the same address, resulting in a communication conflict. Press the button "Address assignment", process the prompt, and confirm the address assignment.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Fig. 50: Dialog box "Maintenance" Sensor adjustment (6) Misadjusted sensors will lead to incorrect measurement values, which in turn will have a negative effect on the process control. CAUTION In this dialog box, the sensors are adjusted in terms of a predefined reference value. • Press the button in order to select the sensor. A selection menu will be displayed. • Enter a reference value and press the button "Adjust". Fig.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation 6.5.1.4 Main window "?" This main window includes the most important information concerning your freeze-dryer: 1 2 3 4 Freeze-dryer type Control system type Serial number Manufacturer contact data 5 Details concerning the software version Fig. 52: Freeze-dryer system information, here for Alpha 2-4 LSCplus In the event of enquiries at the manufacturer, please state the number that is stated here.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Pos: 135 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 60 Bet rieb/ LSCplus/ 060 -00 50- 002 0 G efrie rtr ocknu ng man uell L abo r @ 6 \mo d_1 299 142 7535 61_ 68. docx @ 36 695 @ 3 @ 1 6.5.2 Manual freeze-drying In the manual mode, the user switches manually from one freeze-drying phase to the next. The manual mode is activated by calling up the main window "Manual".
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation The unit will remain in the warm-up phase until you confirm. The transition from "main drying" to "final drying" again takes place automatically. After the completion of the "final drying" phase, the system displays another enquiry with which the freeze-drying process will be terminated. The unit remains in the "run" mode until the enquiry is confirmed. The process can be stopped any time by pressing the "Stop" button.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation condensate water or another medium, the opening time of the drain valve can be pre-selected in this dialog box. • Leakage test (4) This function is optional (see chapter 6.6 - "Optional extensions"). This button displays the parameters that are used for the leakage test. In addition, the test time and the chamber volume must be stated (depending on the accessories that are used).
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Fig. 56: Representation of a freeze-drying process Program sections When the programmer module executes a freeze-drying process, it executes the various sections that were created one after the other until the last section is completed. Within the various sections, the system calculates linear ramps for the temperature and vacuum. These ramps start with the set value of the previous section and end with the set value of the current section.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Conditions for switching to the next section For all of the program sections to be executed automatically, certain switching conditions must be fulfilled at the end of each section. If these conditions are not fulfilled, e.g. due to incorrect set values, a corresponding process message will be displayed. Ice condenser temperature: This value is checked only when the system switches from freezing (loading) to drying. The ice condenser temperature must be -40°C.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Vacuum pump warm-up Apart from the execution of the various sections, the PGMplus programmer module also controls other tasks that need to be performed during the process run, e.g. the automatic activation of the vacuum pump. If the pressure control valve is closed, the PGMplus programmer module shifts the warm-up phase of the vacuum pump to the freezing phase so that it is directly before the first drying section.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Creating a blank program (1) This button is used to call up a blank program template. Only section 1 is pre-defined as the loading section. In this phase, the start conditions of the programs are defined. The room temperature (20°C) is the standard default. If the product is frozen outside the freeze-dryer (double-chamber method), the shelf temperature must be adapted accordingly in section 1.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation • Pressing the button “Diagram” (3) displays the program in the form of a diagram. • Pressing the button “Program name” (2) calls up an input field in which the name of the program can be changed. • The button “Save” (7) can be used in between or at the end of the program creation in order to save the program. • The button “Program” (1) calls up the main window. The system will ask the user whether the program shall be saved if this has not been done yet.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation In contrast to the pressure increase test in the manual mode, repeated pressing of the button in the program mode enables the selection of different variants. • Periodic pressure increase test: The test will be performed periodically during the entire main or final drying phase. The parameters "Duration test" and "Time between tests" apply. The maximum pressure increase is limited to 100%, referring to the actual value at the beginning of the measurement.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Fig. 62: Dialog box "Pressure increase test" in the program mode Option: Comparative pressure measurement (see chapter 6.6 - "Optional extensions") The indication of the measurement value "dp Comp" can be configured in chapter 6.5.1.1 - "Main window "Manual"" under "Actual values of the current process" in the dialog box "Select actual value". NOTE The comparative pressure measurement can be activated or deactivated by pressing the button in the dialog box.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Fig. 63: Dialog box "Comparative pressure measurement" Program templates (2) The PGMplus programmer module offers 14 different program templates that include recipes for all kinds of freeze-drying applications. They are suggestions for orientation purposes and must be adapted to the specific area of application of the freeze-dryer. Following the selection of a program template, a window containing information concerning the drying process will be displayed.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation 6.5.3.2 Editing a program An existing program can be modified as long as it has not been loaded. • Select the program to be edited from the list in the main window “Program”. • Press the button with the program name in order to call up a window that displays the program data. • Perform the desired modifications and save the program (see chapter 6.5.3.1 - "Creating a program"). • Close the dialog box by pressing the “Program” button.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Fig. 65: Dialog box "Load program" • • Press the button "Load" in order to accept the program data. The display switches to the standard user interface. Press the button "Program start" in order to start the freeze-drying process. Fig. 66: The freeze-drying process can be started with the aid of the button “Program start”.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation 6.5.3.5 Pos: 152 /01 0 Univ ersal mod ule/ 74 Deleting a program The PGMplus programmer module offers 32 program storage locations. If they are all occupied, a program must be deleted before a new one can be created. • Select the program to be deleted from the list in the main window "Program". • Press the button with the program name in order to call up a window with the program data. • Press the button “Delete program”.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Pos: 153 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 60 Bet rieb/ LSCplus/ 060 -00 60 Optio nale Erw eite run gen Lab or- -- --- --- -- --- -- --- -- --- --- -- @ 6\ mod _12 991 427 5742 0_6 8.d ocx @ 367 67 @ 2 @ 1 6.6 Optional extensions Comparative pressure measurement see also chapter 6.5.3.1 - "Creating a program", button ">>more" During the sublimation, i.e.
Delta 1-24 LSC& Delta 2-24 LSC& 6 Operation Lyo Control LCplus measuring system The Lyo Control LCplus measuring system can be used to determine the crystallisation state of the product. In the liquid state, the electrical resistance is very low. During the freezing process, the resistance increases. The LyoRx control sensor measures the electrical resistance.
Delta 1-24 LSC& Delta 2-24 LSC& 7 Malfunctions and error correction Pos: 159 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche /07 0 Stö run gen und Fe hlers uche == == === == == === == == === == == == @ 6\ mod _129 913 751 409 1_6 8.doc x @ 3 511 2 @ 1 @ 1 7 Malfunctions and error correction Malfunctions are displayed in the dialog box "Process & equipment messages" (see chapter 7.2 - "Process and error messages").
Delta 1-24 LSC& Delta 2-24 LSC& 7 Malfunctions and error correction Type of error Possible reason Correction The displayed vacuum value is not correct • Incorrect calibration • • The vacuum sensor is soiled (e.g. due to water residues). The vacuum sensor is defective. • • Calibrate the vacuum sensor (see the separate operating instructions of the vacuum sensor). Clean the vacuum sensor. • Check the vacuum display with the aid of a reference device (if available). • see chapter 7.1.2.
Delta 1-24 LSC& Delta 2-24 LSC& 7 Malfunctions and error correction Pos: 166 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0 010 -00 20 Kei n au sreic hend es Vak uum @ 6\ mod _12 991 375 151 07_ 68.d ocx @ 351 36 @ 3 @ 1 7.1.2 Insufficient vacuum The vacuum checks must be carried out when the ice condenser is frozen. NOTE Pos: 167 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.
Delta 1-24 LSC& Delta 2-24 LSC& 7 Malfunctions and error correction Pos: 170 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0 010 -00 20- 0020 Belüf tun gsven til, Me dien abla ufven til @ 6\ mo d_12 991 375 161 22_ 68.d ocx @ 351 52 @ 4 @ 1 7.1.2.2 Pos: 171 /01 0 Univ ersal mod ule/ Aeration valve, media drain valve A malfunction of the aeration valve or the media drain valve may have several causes.
Delta 1-24 LSC& Delta 2-24 LSC& 7 Malfunctions and error correction Pos: 178 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0 010 -00 30 Eisko nde nsat orte mp erat ur Alp ha bis Delta @ 6\ mod _12 991 375 173 10_ 68.d ocx @ 351 76 @ 3 @ 1 7.1.3 Insufficient ice condenser temperature Ensure sufficient ventilation. Do not place any paper, cloth, or similar material behind or under the unit, since otherwise the air circulation will be impaired.
Delta 1-24 LSC& Delta 2-24 LSC& 7 Malfunctions and error correction Pos: 183 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0 030 Kont akt im S ervice fall_L abo r-RVC @ 6\m od_ 129 913 7517 982 _68 .docx @ 35 192 @ 2 @ 1 7.3 Service contact In the event of queries, malfunctions, or spare part enquiries: From Germany • Use the service request form at www.martinchrist.
Delta 1-24 LSC& Delta 2-24 LSC& 8 Maintenance and service Pos: 186 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 80 Wa rtun g u nd In stan dhalt ung/ 080 Wa rtun g un d Ins tan dhalt ung == === == == === == == === == == == == @ 6\m od_ 129 913 772 5871 _68 .docx @ 3 520 1 @ 1 @ 1 8 Maintenance and service The freeze-dryer and the accessories are subject to high mechanical stress. Thorough maintenance performed by the user extends the service life and prevents premature failure.
Delta 1-24 LSC& Delta 2-24 LSC& 8 Maintenance and service Pos: 192 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 80 Wa rtun g u nd In stan dhalt ung/ 080 -00 10- 001 0-0 010 Eiskon dens ato rkam me r @ 6 \mo d_1 299 137 726 840 _68. docx @ 35 225 @ 4 @ 1 8.1.1.1 Pos: 193 /01 0 Univ ersal mod ule/ Ice condenser chamber Before each start-up, ensure that the ice condenser chamber is free from water residues. • Open the media drain valve to drain off any liquid. Then, close the valve.
Delta 1-24 LSC& Delta 2-24 LSC& 8 Maintenance and service Pos: 196 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 80 Wa rtun g u nd In stan dhalt ung/ 080 -00 10- 001 0-0 030 Belüf tung sventil, Me dien ablau fventil @ 6\ mod _12 991 377 275 27_ 68.d ocx @ 352 41 @ 4 @ 1 8.1.1.3 1 2 3 Aeration valve, media drain valve Contaminants such as product residues may lead to an insufficient vacuum. In this case, the aeration valve and the media drain valve must be cleaned.
Delta 1-24 LSC& Delta 2-24 LSC& 8 Maintenance and service Pos: 200 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 80 Wa rtun g u nd In stan dhalt ung/ 080 -00 10- 002 0 Vak uum pu mpe @ 6\ mod _12 991 377 2813 7_6 8.d ocx @ 352 57 @ 3 @ 1 8.1.2 Vacuum pump Please refer to the separate operating manual of the vacuum pump! NOTE The stress of the vacuum pump in conjunction with a freeze-dryer is usually not very high.
Delta 1-24 LSC& Delta 2-24 LSC& 8 Maintenance and service Pos: 204 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 80 Wa rtun g u nd In stan dhalt ung/ 080 -00 10- 004 0 Vak uum mess sond e @ 6\m od_ 129 9137 728 808 _68. docx @ 35 273 @ 3 @ 1 8.1.4 Vacuum sensor Please refer to the separate operating manual of the vacuum sensor! NOTE The vacuum sensor has only a limited service life. Especially carboncontaining substances, e.g. alcoholic compounds, reduce the service life extremely.
Delta 1-24 LSC& Delta 2-24 LSC& 8 Maintenance and service Pos: 208 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 80 Wa rtun g u nd In stan dhalt ung/ 080 -00 20 Desinf ektion von Tr ockn ungsk am mer und Zu beh ör @ 6\ mod _12 9913 772 949 6_6 8.doc x @ 3 528 9 @ 2 @ 1 8.2 Disinfection of the drying chamber and accessories • • • • Use commercially-available disinfectants such as, for example, Incidur , Meliseptol , Sagrotan , Buraton , or Terralin (available at specialised trade).
Delta 1-24 LSC& Delta 2-24 LSC& 8 Maintenance and service Pos: 210 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 80 Wa rtun g u nd In stan dhalt ung/ 080 -00 30 Inst and haltu ngsa rbeit en Labo r-RVC @ 6\ mod_ 129 913 772 984 0_6 8.docx @ 3 529 7 @ 2 @ 1 8.3 Service In the event of service work that requires the removal of the panels, there is a risk of electric shock or mechanical injury. Only qualified specialist personnel is authorised to perform this service work.
Delta 1-24 LSC& Delta 2-24 LSC& 8 Maintenance and service Pos: 212 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 80 Wa rtun g u nd In stan dhalt ung/ 080 -00 40 Rückse ndu ng defek ter Teile @ 6\ mo d_1 2991 377 305 12_ 68.d ocx @ 353 13 @ 2 @ 1 8.4 Return of defective parts Although we exercise great care during the production of our products, it may be necessary to return a unit or accessory to the manufacturer.
Delta 1-24 LSC& Delta 2-24 LSC& 9 Disposal Pos: 215 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 90 Ents org ung/ 090 Ents org ung === == == == === == == === == == === == == === == == === == @ 6\ mod _12 991 379 500 07_ 68.d ocx @ 353 22 @ 1 @ 1 9 Disposal Pos: 216 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 90 Ents org ung/ 090 -00 10 Entso rgu ng d er G efri ertr ockn ungs anlag e @ 6\m od_ 1299 137 950 350 _68. docx @ 35 330 @ 2 @ 1 9.
Delta 1-24 LSC& Delta 2-24 LSC& 10 Technical data Pos: 221 /20 0 Ch rist/3 61 G T-BA L abo r-Pilot (PROJEKTE)/Del ta 1 -24 _2- 24 LSC plus/1 00 Tech nische Date n/1 00 Tec hnisch e Dat en Delt a 1 -24/ 2-2 4 LSCpl us= === == == === == == === == == == === == == == @ 15\ mod _13 554 812 096 09_6 8.
Delta 1-24 LSC& Delta 2-24 LSC& 10 Technical data Connection requirements (without vacuum pump and accessories) Delta 1-24 LSCplus Delta 2-24 LSCplus Electrical connection: 3 x 400 V / 50 Hz 3 x 400 V / 50 Hz Protection class: I I IP protection category according to DIN 60529: 11 11 Nominal power: 0.9 kVA 1.4 kVA Nominal current: 3.5 A 3.
Delta 1-24 LSC& Delta 2-24 LSC& 10 Technical data Special equipment: water cooling system Delta 1-24 LSCplus Delta 2-24 LSCplus Part number: 102302 102303 Cooling water consumption: max. 0.23 m /h max. 0.37 m3/h Heat carried off via the cooling water: 1.9 kW 3.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix Pos: 230 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/1 10 Anh ang /011 0 Anh ang == == === == == == === == == === == == === == == === == == === == == = @ 6\mo d_1 299 138 116 328 _68. docx @ 35 428 @ 1 @ 1 11 Appendix Pos: 231 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/1 10 Anh ang /011 0-0 010 M athe matisc he Zusa mm enh äng e @ 6 \mo d_1 299 138 0925 17_ 68. docx @ 35 388 @ 2 @ 1 11.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix Pos: 233 /20 0 Ch rist/3 61 G T-BA L abo r-Pilot (PROJEKTE)/Del ta 1 -24 _2- 24 LSC plus/1 10 An han g/01 10- 002 0 Kurz bedi enu ngsa nleitu ng Del ta @ 15\ mod _13 5548 259 413 0_6 8.do cx @ 7 811 1 @ 2 @ 1 11.2 Brief operating instructions Functional and operating elements: 1 2 3 LSCplus user interface Vacuum pump Oil mist separator Fig. 74: Front of the freeze-dryer 4 Touchpanel Fig. 75: User interface with touchpanel 96 Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix 5 Contact bolt 6 Pipe connection of the vacuum pump (behind the cover plate) Ice condenser Vacuum sensor 7 8 Fig. 76: Ice condenser chamber 9 Control system on/off 10 Steerable castors Fig. 77: Right side of the freeze-dryer Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix 11 Aeration valve 12 Media drain valve 13 Fixed castors Fig. 78: Left side of the freeze-dryer 14 Vacuum sensor 15 Mains voltage 16 Electrical connection of the vacuum sensor 17 Serial interface 18 Mains cable 19 Heat exchanger of the refrigeration unit (behind the panel) 20 Option: Connection for electrical lifting device 21 Name plate 22 Vacuum connection Fig. 79: Rear side of the freeze-dryer 98 Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix 1 2 Status line Set process values 3 Button "Set values: view/edit" 4 Actual values of the current process Button "Tools" Button "Schematic system diagram" 5 6 7 Button "Process and equipment messages" 8 Button "Operating mode: select/start" Button "Stop" 9 Fig. 80: LSCplus user interface Step-by-step instructions – shelf drying 1 Freeze the sample separately, e.g. in a deep-freeze.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix 11 Start the main drying process either by opening the manual shut-off valve or by waiting for the electromagnetic valve to open. Vacuum is applied to the chamber and the freeze-drying process commences. The vacuum pump always runs with maximum power. With this type of freeze-dryer, the power of the vacuum pump cannot be controlled. NOTE 12 The operating panel displays the vacuum, the ice condenser temperature, and the current operating mode.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix 4 Ensure that all of the valves are closed. 5 Let the vacuum pump warm up 20 to 30 minutes before the freezedrying processes commences. 6 Connect a frozen sample to a valve. After the pressure has fallen below 1.030 mbar, a frozen sample can be connected to a valve. The next frozen sample cannot be connected to another valve until the pressure is again lower than 1.030 bar. CAUTION The vacuum pump always runs with maximum power.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix Pos: 235 /20 0 Ch rist/3 61 G T-BA L abo r-Pilot (PROJEKTE)/Del ta 1 -24 _2- 24 LSC plus/1 10 An han g/01 10- 030 Konf or mitätse rklä run g @ 1 5\m od_ 135 548 2595 614 _68 .docx @ 78 117 @ 2 @ 1 11.3 Pos: 236 /01 0 Univ ersal mod ule/ 102 EC declaration of conformity Sei tenwe chsel @ 0\ mod _12 021 1624 431 2_0 .docx @ 1 05 @ @ 1 Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix Pos: 237 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/1 10 Anh ang /011 0-0 040 Spa nnu ngsriss - u nd Ch emik alienb estä ndigk eit Plexiglas @ 1 4\m od_1 352 961 626 395 _68. docx @ 75 126 @ 2 @ 1 11.4 Resistance to stress cracking an chemical influences "Plexiglas" Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix 104 Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix 106 Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix 108 Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix 110 Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix 112 Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix 114 Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix 116 Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix 118 Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix Pos: 238 /01 0 Univ ersal mod ule/ Sei tenwe chsel @ 0\ mod _12 021 1624 431 2_0 .docx @ 1 05 @ @ 1 Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 11 Appendix Pos: 239 /01 0 Univ ersal mod ule/9 00 Gl ossa r GT La bor == === == == === == == === == == == === == == === == == === == @ 3\ mod _12 849 666 568 22_ 68.d ocx @ 247 84 @ 1 @ 1 120 Version 04/2013, Rev. 1.
Delta 1-24 LSC& Delta 2-24 LSC& 12 Glossary 12 Glossary Comparative pressure measurement During the sublimation, i.e. when the concentration of water vapour molecules is rather high in the atmosphere, the value that is provided by the gas-type dependent vacuum sensor of the "Pirani" type (e.g. Thyracont VSP 62/63) deviates from the value that is provided by a capacitive vacuum sensor (e.g. MKS 722B).
Delta 1-24 LSC& Delta 2-24 LSC& 12 Glossary Double-chamber method Drying on shelves outside the ice condenser chamber is referred to as a double-chamber system. The advantage compared to the singlechamber method is the considerably higher product capacity. In addition, the product chamber can be isolated from the ice condenser chamber by an intermediate valve in order to perform a so-called pressure increase test for determining the end of the drying process.
Delta 1-24 LSC& Delta 2-24 LSC& Index 13 Index A Access protection....................................... 53 Accessories ................................... 11, 27, 87 Accident prevention ................................... 25 Actual values of the current process .......... 42 Administration ............................................ 51 Aeration ..................................................... 62 Aeration valve .......................... 14, 34, 93, 98 Aeration valve, cleaning and care ............
Delta 1-24 LSC& Delta 2-24 LSC& Index Disposal of the freeze-dryer ....................... 91 Disposal of the packaging .......................... 91 Double-chamber method.......................... 122 Double-chamber method (Process B) ........ 50 Double-chamber method (Process B) – LyoCube ................................................. 50 Drying chamber.......................................... 84 E EC declaration of conformity .............. 11, 102 Editing a program...................................
Delta 1-24 LSC& Delta 2-24 LSC& Index M Main drying ................................................ 20 Main window "?"......................................... 60 Main window "Manual" ............................... 40 Main window "Options" .............................. 49 Main window "Program" ............................. 48 Mains cable.......................................... 14, 98 Mains fuse ................................................. 93 Mains voltage.......................................
Delta 1-24 LSC& Delta 2-24 LSC& Index Protective gloves........................................ 27 R Radioactive substances ....................... 28, 83 Rated current ............................................. 15 Reference designator............. 44, 46, 56, 122 Refrigerant ................................................. 93 Refrigerant data ......................................... 15 Registered office ........................................ 15 Remaining hazards ....................................
Delta 1-24 LSC& Delta 2-24 LSC& Index Vapour pressure curve above ice ............... 17 Vapour pressure curve for ice and water.... 42 Ventilation ............................................ 33, 81 Viewing or editing the set values ................ 62 Warranty and liability.................................. 11 Water cooling system................................. 94 Weight ................................................. 30, 93 Wireless shelf technology ..................