7.1.2 Insufficient vacuum
NOTE
The vacuum checks must be carried out when the ice condenser is frozen.
Pos: 168 /01 0 Univ ersal module/ Lee rzeile @ 0\ mo d_12021 162445 00_ 0.docx @ 1 14 @ @ 1
Pos: 169 /200 Ch ris t/360 G T-BA L abo r-Pil ot (STA NDA RDMODU LE )/0 70 Stö rungen und Fe hlers uche /07 0-0010 -00 20-0010 K leinfl ans chv erbi ndunge n @ 6\mo d_1 299 137515 528 _68. docx @ 35144 @ 4 @ 1
7.1.2.1 Small flange connections
Leakages are often due to improper small flange connections between the
various components and hose connections or to leakages in the valves.
• Loosen the connection and place the centring ring (with sealing ring
inside) in a centred manner between the flange connections.
• Seal the connection with the clamping ring by tightening the wing nut.
• Ensure that the centring ring neither slips out of place nor gets jammed.
Fig. 65: Small flange and centring ring Fig. 66: Small flange with centring ring
and small flange
Fig. 67: Attaching the clamping ring Fig. 68: Tightened clamping ring