7 Malfunctions and error correction
Version 11/2006, Rev. 2.0 of 02/12/2014 • sb
55
Translation of the original operating manual
Error message Possible cause Remedy
Ice condenser temperature 20°C
The ice condenser temperature rises to
a non-permissible value during the
freeze-drying process. The pressure
control valve will close automatically in
order to prevent the vacuum pump from
being damaged.
Insufficient cooling. Observe the behaviour of the
refrigeration system. If
necessary, inform the service
department (see chapter 7.3 -
"Service contact").
Defective ice condenser temperature
sensor
The temperature sensor is not
connected or it is defective.
Inform the service department
(see chapter 7.3 - "Service
contact").
Defective temperature sensor of the
defrosting heater
The temperature sensor is not
connected or it is defective.
Inform the service department
(see chapter 7.3 - "Service
contact").
Mains power failure (see chapter 7.1.1
- "Power failure")
The power supply has been
interrupted.
The mains power connector is
not connected.
The fuses have tripped.
The mains power switch has
been switched off.
Check the mains fuse.
Connect the mains power
connector firmly.
Check the customer-provided
fuses.
Switch the mains power switch
on.
Default settings loaded
All of the settings have been reset to the
delivery state of the freeze-dryer.
Error in the parameter memory.
Manual execution in the service
menu.
No measure required.
IO communication error (0x20) Error in the LDplus IO module. Inform the service department
(see chapter 7.3 - "Service
contact").
Pos: 159 /01 0 Uni v ersal mod ule/L eerz ei le @ 0\m od_ 120 211624 450 0_0. doc x @ 11 4 @ @ 1
Pos: 160 /01 0 Uni v ersal mod ule/L eerz ei le @ 0\m od_ 120 211624 450 0_0. doc x @ 11 4 @ @ 1
Pos: 161 /20 0 Chri st/3 60 GT-B A L abor-P ilot (STAN DAR DMODU LE )/0 70 Stö rungen und Fe hlers uche /07 0-0 030 Kont akt i m S ervice fall_L abo r-RV C @ 25\ mod _14 0498406 875 8_6 8.do cx @ 1 84279 @ 2 @ 1
7.3 Service contact
In the event of queries, malfunctions, or spare part enquiries:
From Germany:
Contact
Martin Christ Gefriertrocknungsanlagen GmbH
An der Unteren Söse 50
37520 Osterode (Germany)
Tel. +49 (0) 55 22 / 50 07-44 44
E-mail: support.lab@martinchrist.de
Outside Germany:
Contact our agency in your country. All agencies are listed at
www.martinchrist.de [Sales Partners]
NOTE
If you would like to utilise our after-sales-service, please state the type of
your freeze-dryer and its serial number.
Pos: 162 /01 0 Uni v ersal mod ule/A bsc hnit tsw echs el @ 0\ mod _12021 245 140 62_ 0.docx @ 418 @ @ 1
Pos: 163 /01 0 Uni v ersal mod ule/S eit enwec hsel @ 0\m od_ 120 211 624 4312 _0. docx @ 10 5 @ @ 1