Instruction Manual

Freeze-dryer Alpha 1-2 LDplus
7 Malfunctions and error correction
Version 11/2006, Rev. 2.0 of 02/12/2014 • sb
51
Translation of the original operating manual
Fig. 41: Small flange and centring ring Fig. 42: Small flange with centring ring
and small flange
Fig. 43: Attaching the clamping ring Fig. 44: Tightened clamping ring
Pos: 147 /01 0 Uni v ersal mod ule/L eerz ei le @ 0\m od_ 120 211624 450 0_0. doc x @ 11 4 @ @ 1
Pos: 148 /20 0 Chri st/3 60 GT-BA Labo r-Pi lot (STAND ARD MODU LE)/0 70 Stö run gen und Fe hlers uche /070-0 010 -00 20-0020 B elüf tun gs- und Medi ena blaufv entil L Dpl us @ 31\ mo d_1 4084386 391 12_ 68.d ocx @ 209 718 @ 4 @ 1
7.1.2.2 Aeration and media drain valve
A malfunction of the aeration and media drain valve may have several
causes. One potential source are contaminants such as product residues
within the valve.
Switch the freeze-dryer off and disconnect the mains plug.
Clean the valve (see chapter 8.1.1.3 - "Aeration valve, media drain
valve").
Put the freeze-dryer into operation again.
If there is still a leakage, the freeze-dryer must be checked by qualified
specialist personnel (see chapter 7.3 - "Service contact").
Pos: 149 /01 0 Uni v ersal mod ule/L eerz ei le @ 0\m od_ 120 211624 450 0_0. doc x @ 11 4 @ @ 1