7 Malfunctions and error correction
50
Version 11/2006, Rev. 2.0 of 02/12/2014 • sb
Translation of the original operating manual
Pos: 139 /200 Ch rist /360 G T-BA L abo r-Pilot (STA NDA RDMODU LE )/070 S törun gen und Fe hlers uche /070 Stö rungen und Fe hlers uche == ===== ==== ===== ===== ==== == @ 25\ mo d_14049 840 560 39_68. docx @ 184 111 @ 1 @ 1
7 Malfunctions and error correction
Malfunctions are displayed in the dialog box "Process & equipment
messages" (see chapter 6.5.3.3 - "Process and equipment information").
An acoustic signal sounds when an error message is generated.
Eliminate the source of the problem (see the following chapter).
Acknowledge the error message.
Pos: 140 /01 0 Uni v ersal mod ule/L eerz ei le @ 0\m od_ 120 211624 450 0_0. doc x @ 11 4 @ @ 1
Pos: 141 /20 0 Chri st/3 60 GT-B A L abor-P ilot (STAN DAR DMODU LE )/0 70 Stö rungen und Fe hlers uche /07 0-0 010 Al lgem eine St öru ngen LD pl us @ 31\ mod _140 843825 641 5_6 8.doc x @ 2 09704 @ 2 @ 1
7.1 General malfunctions
Pos: 142 /20 0 Chri st/3 60 GT-BA Labo r-Pi lot (STAND ARD MODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0010 -00 10 Netzau sfal l @ 25\ mod _14 049840 5896 4_6 8.d ocx @ 1841 53 @ 3 @ 1
7.1.1 Power failure
The control system continues with the process after a power failure. The
preselected conditions remain saved even during a process run.
In the event of a power failure in the drying phase, the batch may become
unusable. Whether the batch can be saved or not depends on the drying
phase in which the product was when the power failure occurred.
In the final drying phase, the product has reached a residual moisture
content of approx. 5%. Below this value, the product is generally not
damaged even if the power failure lasts for a longer period of time.
If the product is in the main drying phase, we recommend aerating the
unit, removing the product, and storing it in a deep-freeze. The
defrosted condensate must be drained off prior to the next start.
Pos: 143 /01 0 Uni v ersal mod ule/L eerz ei le @ 0\m od_ 120 211624 450 0_0. doc x @ 11 4 @ @ 1
Pos: 144 /20 0 Ch rist/3 60 G T-BA Labo r-P ilot (STA NDA RDMODU LE )/070 S tö rungen und Fehl ers uche /07 0-0010 -00 20 Kei n au sreic hend es V ak uum @ 2 5\mo d_1 404 984060 039 _68. docx @ 18 416 7 @ 3 @ 1
7.1.2 Insufficient vacuum
NOTE
The vacuum checks must be carried out when the ice condenser is frozen.
Pos: 145 /01 0 Uni v ersal mod ule/L eerz ei le @ 0\m od_ 120 211624 450 0_0. doc x @ 11 4 @ @ 1
Pos: 146 /20 0 Chri st/3 60 GT-BA Labo r-Pi lot (STAND ARD MODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0010 -00 20-0010 K leinf l anschv er bindu ngen @ 25\m od_140 498 406 1086_68 .docx @ 1 841 81 @ 4 @ 1
7.1.2.1 Small flange connections
Leakages are often due to improper small flange connections between the
various components and hose connections or to leakages in the valves.
Loosen the connection and place the centring ring (with sealing ring
inside) in a centred manner between the flange connections.
Seal the connection with the clamping ring by tightening the wing nut.
Ensure that the centring ring neither slips out of place nor gets jammed.